MACH3
August 2008
0540.714
    Technical user manual

 
 


Introduction
 


 

The MACH3 system consists of a four circle goniometer (Omega, Kappa, Phi to position the crystal and Theta to position the detector) and an X-ray source. The detector measures X-rays diffracted from a crystal.

X-rays can be generated by a sealed tube which is part of the goniometer and a HV generator or by a stand alone rotating anode generator. The X-ray optics consists of a ß filter or a monochromator to suppress the ß- and brehms radiation and a collimator to get a small beam. Safety switches close the shutter or switch off the high voltage in unsafe conditions. In case of a rotating anode generator the shutter and some safeties are handled by the generator. The optics at the detector side consists of a collimator, replaceable slits, a dial unit with computer selectable slits and an attenuator to attenuate high intensities.

The axes are decoded by precision optical encoders and are driven by DC motors, that permit accurate, fast and smooth positioning with low power consumption. The goniometer is driven by an electronic interface, which has a processor, drivers and detector circuit. The system is run by software through the main program on the computer. A selected number of functions, which are needed in the vicinity of the goniometer, can be executed by a pocket terminal. The goniometer is surrounded by a radiation enclosure, which prevents scattering of radiation into the room.


Safety instructions Alterations, other than mentioned in this manual, can obstruct the safety precautions and are therefore not recommended. This includes adjustments to parts which are sealed. Parts may only be changed by identical spare parts.

X-ray safety

The equipment generates open X-rays, which can cause to humans hazards when safety rules are not followed:

Beryllium

The detector and X-ray tube contain windows of beryllium. In normal use these area's cannot be reached, but during maintenance, care must be taken not to touch the window. When by accident the Beryllium is touched, wash with soap and water.

Fumes and dust from beryllium metal can be hazardous if inhaled. During use, corrosion of the beryllium window may occur, but it should not be cut, machined or otherwise removed. Tube and detector disposal must comply with the national and local regulations governing beryllium.
 
 


Specifications
Angle, DX
Phi  ±180°
Kappa ±175°
Omega ±210°
Theta, with sealed tube -49°, +77° 
Chi 0° - 99.85°
Speed
Phi  0.0025 - 16.48 °/min 
Kappa, Omega, 2Theta 0.0017 - 16.48 °/min
Search, all axes  max. 197.8 °/min
Accuracy
Phi  ±0.015°
Kappa, Omega, 2Theta ±0.010°
Reproducibility ±0.005°
Resolution 0.0003°
Sphere of confusion R = 5 µm
--
Load, Phi max. 1 kg
Load, Detector max. 6 kg
Microscope 75 x, grid 0.036 mm/div
Second microscope (option)  125 x

Anti-collision

X-ray optics
X-ray source
Power maximum 60 kV, 300 mA
Crystal - focal point 216.5 mm
Beta filters
Open hole  beta/alpha = 13%
Cu 0.020 mm nickel, beta/alpha = 0.2%
Mo 0.108 mm zirconium, beta/alpha = 0.2%
Fe 0.016 mm manganese, beta/alpha = 0.2%
-
Monochromator, sealed tube Graphite, 12 x 6 mm, 0 0 2 reflection, mosaic spread approx. 0.4°
Collimators, sealed tube
    Fine focus
0.4 mm (1 ring), 0.8 mm (2 ring), 1.3 mm (3 ring)
    Normal focus
0.8 mm (1 ring), 1.3 mm (2 ring) 
Dial, software control
0,2 mm slits +45° (Pos), -45° (Neg), vertical, horizontal
Aperture horizontal 1.3 - 5.9 mm and open hole 9 mm.
Reproducibility slits ±0.01 mm.
Crystal - slit 173 mm
Aperture vertical (manual)  1 mm, 2 mm, 3 mm, 4 mm, 6 mm.
Attenuator Cu : 0.06 mm Nickel, typical attenuation, 14x
Mo: 0.26 mm Zircon, typical attenuation, 14x
Beamstop Fine focus : 8 mm outside, inside Ø5.7 x 4 mm lead.
Normal focus : 12 mm outside, inside Ø9.1 x 4 mm lead.
Collimators detector 2 mm, 5 mm.
Count rate  <500 kcps
Accuracy  <1%
Scintillator  NaI(T1)
Absorption  Cu - Ag: 100%
Dead time  Automatic compensation
Extension detector arm (option)  Crystal - slit : 368 mm

Goniometer heads
 
Height  49.0 mm
Distance ring 14.96 mm
ECON head X, Y: ±1.5 mm Z: 8 mm
XYZ head (option) X, Y: ±2.5 mm Z: 8 mm
Eucentric head (option)  X, Y: ±2.5 mm Z: 3 mm angle: ±15° 

The shutter does not open, or is closed when: The X-ray high voltage is not switched on, or is switched off when: Safety override

The system is supplied with two overrides, which are only allowed to be used during technical maintenance (see safety rules).

Safety with rotating anode generator FR591

With a X-ray generator all safeties are handled by the generator, except the shutter safeties: collimator and beam stop.

Safety signals

With all lamps, excluded the green shutter-closed lamp, a defect lamp is detected and activates the safety.

Safety with power off

Safety demands that circuits are automatically in a safe position when they are out of order. So during power off, the goniometer Interface and the X-ray high voltage are switched off by the safeties.

Data-line to Host  RS232: 300-19200 bd (auto-baud), 8 databits, 1 stopbit, no parity. Galvanic separated.
Crystal illumination  Supply 24 V, max. 80 mA, 2 W (not used).
X-ray / HV generator  Hardware signals for shutter- and safety actions and a RS232 for software communication. 
Environment Indoor use
Altitude  Up to 2000 m
Temperature, operating 20-22°C (within specification)
Relative humidity Maximum 80% if <31°C, decrease linearly to 50% at 40°C
Pollution degree 2, according to IEC 664 (normal clean environment).
Protection IP 20, according to IEC 529 (limited protection to water).
Gasses No explosive gasses are allowed.
--
Power supply Nominal 100V-120 V, 200V-240 V. max. 150 VA, single phase
Fluctuations ±10% 
Frequency 47-63 Hz. 100V-110 V: 57-63 Hz.
Installation category II, according to IEC 664.
Goniometer 146 kg
Interface 8.5 kg


Installation

Position

Power Interface

The Interface is mounted in the cabinet or placed on the table top. The Interface must not be placed within 50 cm of the X-ray tube, to prevent electrical problems due to spilt cooling water. Ventilation holes must not be covered and no insulating or heavy object must be placed on the equipment.

Safety ground



Connections and controls
 
 


Operation Operation of the MACH3 is executed through the computer program: see the software manual. A limited number of operations can be executed by the pocket- terminal.
 

Start-up procedure

Pocket terminal
 
F1 - F4 Function or submenu
^, v Scroll through a setting
* Specials
Shutter  Only active in Detector menu
<, >  Position axis
STOP  Stops all motors and closes the shutter.
Axes control  Positioning of the four axes. Position in degrees
View positions Gives a choice of crystal align positions 
Detector Displays the contrite of the detector. Open/close shutter. It is used for technical alignment of the goniometer
- * - Special functions for maintenance
Other messages

 
 
Axes control
The +###.### *  Theta position
Phi +###.### *  Phi position
Ome +###.### *  Omega position
Kap +###.### *  Kappa position

character * denotes : blank = motor stop, R = motor Running
 
Submenu Axes control 
Theta +###.### Position with < and >
Speed # Speed with ^ and v, blank: variable, from slow to fast 
2000 °/min  200 °/min  20 °/min 
8 1000 °/min  5 100 °/min  2 10 °/min 
7 500 °/min  4 50 °/min  1 5 °/min 

 
 
View positions 
View 0 Positions to align the crystal View 0, 1 : Position in front of the goniometer
View 2, 3 : Position at the side of the goniometer
The choice between 0, 1 or 2, 3 is made in the host program in the goniometer constants
View 1
Phi +90  Rotates Phi 90°. Phi also rotates with < and >
Zero All axes 0°

 
 
Detector
###### cnt/s Intensity, measured by the detector
Smooth ## s Average time: choose with ^ and v, default 1
Dial aper #.# mm Dial: shows slits or aperture.
Attenuator OFF Attenuator on/off with F4.

 
 
Specials : *
Omega +90 degr Rotate Omega +90° 
Omega -90 degr Rotate Omega -90°
Release motors Power off motors 
Park Position for quick zero-find after power off (will also be executed automatic during exit of the program).

 
 
Other messages
Connecting to host computer Tries to connect to the host program during start-up
Zeroing procedure ongoing Goniometer is running slowly to zero to determine the zero reference position during start-up.
< Theta > Collision. See chapter Trouble shooting. 


Maintenance Cleaning

For cleaning use a damp cloth, moistened with water or isopropyl alcohol. No cleaning agent may enter the goniometer.


Tools


Safety test
  1. Switch off the X-ray high voltage
  2. Open the shutter. Check both shutter-open lamps.
  3. Check if the shutter will not open after activating the next unsafe conditions. The shutter must remain closed after reestablishment of the safety (except the shutter lamps):
  4. Check the flashing Open-beam lamp when the radiation enclosure safety is override by the key-switch.
X-ray high voltage safety, with the FR590
  1. Switch on the high voltage to minimum: 20 kV 5 mA
  2. Check the X-ray on lamps.
  3. Check if the high voltage falls off when:
X-ray high voltage safety, with the FR591
  1. Switch on the high voltage on minimum: 20 kV 5 mA.
  2. Check if the high voltage falls off when:
  3. See safety tests in FR591 service manual.

X-ray optics alignment with a sealed tube The measurement is done in a less safe condition : without beamstop and the enclosure safety override. Safe operation is possible when executed by a qualified person and precautions have been taken (see safety instructions)
Measurement set-up
  1. Place the 1-ring collimator.
  2. Place the absorber on the collimator (take care: Mo and Cu use different absorbers).
  3. Position on Zero, aperture on open hole, no beta filter.
  4. Remove the beamstop and place the safety switch override. take care, the primary beam is measured, be sure Theta is always 0º or use a second beamstop (X-ray shield) behind the goniometer (the PVC windows are not sufficient to stop the primary beam), see X-ray safety instructions.
  5. X-ray generator:
  6. For measuring the intensity use the pocket terminal : detector.
Adjustment
  1. Adjust Tube 2Theta to 0°.
  2. Dismount the monochromator and mount the X-ray port (two Allen screws {2.5 mm} at the bottom, which can be reached by two holes {not the middle one} under the black monochromator holder). . During replacement the power of the X-ray generator must be off.
  3. Adjust tube 2Theta = 0° , and take-off = 3.0° (for fine focus 0.4 x 8 mm, for normal 1 x 10 mm focus : 6.0°).
  4. Adjust on approx. maximum intensity with tube vertical (Allen 4 mm) and tube horizontal (Allen 3 mm) adjustment.
  5. Position slit horizontal and adjust on maximum intensity with tube vertical adjustment.
  6. Position slit vertical and adjust on maximum intensity with tube horizontal adjustment.
  7. If needed repeat both and lock the tube vertical adjustment (Allen 5 mm).
  8. Remove the X-ray port. Switch off the power of the X-ray generator during replacement.
  9. Mount the monochromator and switch on the power of the X-ray generator
  10. Adjust Tube 2Theta (Mo: 12.2°, Cu: 26.6°) and the Monochromator 2Theta on the anode target material used.
  11. Position on slit horizontal and adjust on maximum intensity, first with monochromator 2Theta, then tube 2Theta and monochromator height.
  12. Position on slit vertical and adjust on maximum intensity with tube horizontal adjustment.
  13. Repeat both several times. Lock the monochromator height and tube 2Theta.
  14. Mount the test crystal (Ammonium bitartrate), place the beamstop and remove the absorber, use the 2-ring collimator, set X-ray generator: 50 kV, 20 mA (Mo and Cu).
  15. Input top reflexes, or search for reflections, if the matrix is unknown.
  16. Run program Align and adjust till the output is <0.01 (mm). This adjustment will lower the intensity somewhat :
  17. Lock monochromator 2theta.
  18. Check the intensity with the Effic measurement
    Adjustment points

 


EFFIC: intensity of the primary beam

The effic measurement is a relative measurement of the intensity of the primary beam, which measures the total efficiency of the X-ray: tube and optics. When the value is too low it is mostly caused by the tube (degradation) or the X-ray alignment. The measurement is done during installation of the system and later measurements can be compared with this value. Take into account that because of tolerance in collimator and absorber (material thickness) the values cannot be accurately compared with other systems.

Measurement:


Reset

When the goniometer program halts, it can be reset by the button on the front of the interface. A reset is also possible by switching off the power but this will also switch off the X-ray high voltage. See also the start-up sequence.


Detector HV adjustment

Because the detector amplifier tube will degrade in time the HV (which controls the gain) should be adjusted regularly (every 1-2 year). The detector has two HV settings:


User replaceable parts
 
X-ray tube replacement
Monochromator
  1. First switch off the power of the X-ray generator.
  2. The monochromator is fixed with screws from below the collimator stand.
  3. First adjust the Tube-2Theta before mounting.
  4. With the X-ray port: first adjust Tube-2Theta before mounting.
  5. When not mounted in the right position the shutter will not open.
  6. After changing: realign the X-ray optics.
When mounting the collimator, the key must match the slot. When not mounted in the right position the shutter does not open. After changing, check if it is in good alignment with the collimator.
  • Attach the microscope to the base of the goniometer. Make sure that it is perfectly vertical.
  • Mount the test crystal and centres it, using the fixed microscope.
  • Adjust the position of the second microscope as to get the test crystal in the centre of the reticule, with adjustments height and rotate. Lock the adjustments.
  • The microscope must be removed from the stand before operating the goniometer. The stand does not interfere with any axis movement and can be left to the base of the goniometer. 
  • Position the goniometer with the POLA command. The polaroid cassette may only be used in this position. 
  • Mount the Polaroid cassette on the Omega axis. 
  • Loading the film :
    • Set the LOAD/PROCESS lever to LOAD. 
    • Insert the film pack firmly with 'this side toward lens' away from the crystal. 
    • Pull out the film pack until it stops. Pull the EXPOSE knob out. The film is ready to expose. 
Axial polaroid cassette (option)
Replaceable parts Fuses must only be replaced with the power cable disconnected.
 
Interface, primary 4 A T, 4 x 20 mm, On the outside of the Interface. 
Goniometer
    24 V
1.6 A T, 4 x 20 mm, F2 on the internal connection board
    Crystal Ill. supply
200 mA T, 4 x 20 mm, F1 on the internal connection board
24 V, 50 mA, 1,2 W, BA9s. Long life type. part number 2603.335 (order in lots of 10). It is not allowed to install a LED lamp, because a defect lamp is not detected by the Safety interlocks.
 
O-ring seals
Test crystal
Ammonium bitartrate: (the crystal is different from a KappaCCD system : larger size and oriented). Change: programmable IC for safety (KappaCCD/MACH3 sold until 2002)

When the safety doors open the shutter will close for safety. When at that moment the shutter does not close completely the HV will not be switched off as under all other circumstances will happen. To prevent this problem an electronic component (IC) must be replaced. This is not difficult and can be done by an electronic engineer. It is not needed for KappaCCD/CCD on a rotation anode generator.


Ceramic tubes

When using Siemens ceramic tubes with the Nonius HV connector an isolation tube is needed. This must be placed as shown below:


Accessories box


Troubleshooting Start-up and communication problems
Collision and maximum angle switch

The anti-collision software prevents the goniometer from colliding with the collimator holders or the axes to run past their limits. When a collision takes place because of unexpected circumstances, like a technical error, it is sometimes possible to return:

  1. The pocket terminal shows a special menu: choose the axis you want to position with F1-4.
  2. First press the overrule button on the goniometer (near the connectors).
  3. Then press simultaneously the pocket terminal < or >. Take care not to run the goniometer further into collision.
  4. As soon as the collision switch is released, the goniometer starts to zero-find.

Zero finding Check the dummy plug on the goniometer connector 'Limit switch'.



Software error messages which can be caused by technical reasons are:
 
'Crash switch detected'  The goniometer has touched a collision switch on a collimator holder. Try to run out of collision. Check why the system came into collision, because the software should prevent this
'Generator problem' The X-ray high voltage is off. The generator will show the cause. Possible reasons in the goniometer are: 
  • The shutter does not close (also indicated by the shutter lamp). When this happens the shutter must be repaired.
  • The high voltage cable is not connected to the X-ray tube. This is indicated by a link, attached to the cable and connected to the X-ray on lamp.
  • The X-ray on lamp is defect. 
  • 'Shutter problem' The shutter does not open. This can be caused by: 
  • The radiation enclosure is not closed.
  • The collimator is not in the right position. 
  • The monochromator or X-ray port is not in the right position.
  • The beamstop is not mounted. 
  • The shutter-open lamp is defect. 
  • When the radiation enclosure safety is override with the key-switch:
  • Beam open lamp (with radiation symbol) is defect. 

  • X-ray intensity problem
    Shutter does not open
    Manufacturer and Service

    BRUKER AXS B.V., P.O. Box 811, 2600 AV Delft, The Netherlands, Tel: 31 15 2152400 Fax: 31 15 2152500.

    Service: see our web site: www.bruker-axs.com