| MACH3 |
 |
August 2008
0540.714
Introduction
The MACH3 system consists of a four circle goniometer (Omega, Kappa,
Phi to position the crystal and Theta to position the detector) and an
X-ray source. The detector measures X-rays diffracted from a crystal.
X-rays can be generated by a sealed tube which is part of the goniometer
and a HV generator or by a stand alone rotating anode generator. The X-ray
optics consists of a ß filter or a monochromator to suppress the
ß- and brehms radiation and a collimator to get a small beam. Safety
switches close the shutter or switch off the high voltage in unsafe conditions.
In case of a rotating anode generator the shutter and some safeties are
handled by the generator. The optics at the detector side consists of a
collimator, replaceable slits, a dial unit with computer selectable slits
and an attenuator to attenuate high intensities.
The axes are decoded by precision optical encoders and are driven by
DC motors, that permit accurate, fast and smooth positioning with low power
consumption. The goniometer is driven by an electronic interface, which
has a processor, drivers and detector circuit. The system is run by software
through the main program on the computer. A selected number of functions,
which are needed in the vicinity of the goniometer, can be executed by
a pocket terminal. The goniometer is surrounded by a radiation enclosure,
which prevents scattering of radiation into the room.
Safety
instructions
Alterations, other than mentioned in this manual, can obstruct the safety
precautions and are therefore not recommended. This includes adjustments
to parts which are sealed. Parts may only be changed by identical spare
parts.
X-ray
safety
The equipment generates open X-rays, which
can cause to humans hazards when safety rules are not followed:
-
The equipment must be operated under the responsibility of an X-ray safety
officer, as stated by local regulations.
-
The equipment must be operated, serviced and maintained by qualified personnel
who are trained with the hazards of, and precaution against X-ray exposure.
-
The equipment contains safety circuits which should not be overridden,
other than supplied: the enclosure key and beamstop override vane. These
overrides may only be used for service and maintenance, i.e. X-ray alignment.
-
The enclosure key and beamstop override vane should not be kept with the
instrument but under the custody of the X-ray safety officer. This also
applies to the X-ray- or shutter safety keys of the X-ray generator in
case the X-ray port is not used or X-ray is not allowed to be used.
-
When using the beamstop override vane (for alignment only) the system must
not be left unattended. Theta always at 0°, safety door closed, low
kV/mA setting and a second lead beamstop.
-
Safety circuits must be tested by service personnel or the safety officer
at least every 12 months, as described in this manual
-
The collimator must always be installed.
Beryllium
The detector and X-ray tube contain windows of beryllium. In normal
use these area's cannot be reached, but during maintenance, care must be
taken not to touch the window. When by accident the Beryllium is touched,
wash with soap and water.
Fumes and dust from beryllium metal can be hazardous if inhaled. During
use, corrosion of the beryllium window may occur, but it should not be
cut, machined or otherwise removed. Tube and detector disposal must comply
with the national and local regulations governing beryllium.
Specifications
| Angle, DX |
|
| Phi |
±180° |
| Kappa |
±175° |
| Omega |
±210° |
| Theta, with sealed tube |
-49°, +77° |
| Chi |
0° - 99.85° |
|
|
| Speed |
|
| Phi |
0.0025 - 16.48 °/min |
| Kappa, Omega, 2Theta |
0.0017 - 16.48 °/min |
| Search, all axes |
max. 197.8 °/min |
|
|
| Accuracy |
|
| Phi |
±0.015° |
| Kappa, Omega, 2Theta |
±0.010° |
| Reproducibility |
±0.005° |
| Resolution |
0.0003° |
| Sphere of confusion |
R = 5 µm |
| -- |
|
| Load, Phi |
max. 1 kg |
| Load, Detector |
max. 6 kg |
| Microscope |
75 x, grid 0.036 mm/div |
| Second microscope (option) |
125 x |
Anti-collision
-
Automatic anti-collision software, also when using the pocket terminal..
-
Axis maximum position switches. Turning back is possible.
-
Collision switches on both collimator holders.
X-ray optics
| X-ray source |
|
| Power |
maximum 60 kV, 300 mA |
| Crystal - focal point |
216.5 mm |
|
|
| Beta filters |
|
| Open hole |
beta/alpha = 13% |
| Cu |
0.020 mm nickel, beta/alpha = 0.2% |
| Mo |
0.108 mm zirconium, beta/alpha = 0.2% |
| Fe |
0.016 mm manganese, beta/alpha = 0.2% |
| - |
|
| Monochromator, sealed tube |
Graphite, 12 x 6 mm, 0 0 2 reflection, mosaic spread approx. 0.4° |
| Collimators, sealed tube |
|
|
|
0.4 mm (1 ring), 0.8 mm (2 ring), 1.3 mm (3 ring) |
|
|
0.8 mm (1 ring), 1.3 mm (2 ring) |
|
|
| Dial, software control |
|
| 0,2 mm slits |
+45° (Pos), -45° (Neg), vertical, horizontal |
| Aperture horizontal |
1.3 - 5.9 mm and open hole 9 mm. |
| Reproducibility slits |
±0.01 mm. |
| Crystal - slit |
173 mm |
| Aperture vertical (manual) |
1 mm, 2 mm, 3 mm, 4 mm, 6 mm. |
|
|
| Attenuator |
Cu : 0.06 mm Nickel, typical attenuation, 14x
Mo: 0.26 mm Zircon, typical attenuation, 14x |
| Beamstop |
Fine focus : 8 mm outside, inside Ø5.7 x 4 mm lead.
Normal focus : 12 mm outside, inside Ø9.1 x 4 mm lead. |
| Collimators detector |
2 mm, 5 mm. |
| Count rate |
<500 kcps |
| Accuracy |
<1% |
| Scintillator |
NaI(T1) |
| Absorption |
Cu - Ag: 100% |
| Dead time |
Automatic compensation |
|
|
| Extension detector arm (option) |
Crystal - slit : 368 mm |
Goniometer heads
| Height |
49.0 mm |
| Distance ring |
14.96 mm |
| ECON head |
X, Y: ±1.5 mm Z: 8 mm |
| XYZ head (option) |
X, Y: ±2.5 mm Z: 8 mm |
| Eucentric head (option) |
X, Y: ±2.5 mm Z: 3 mm angle: ±15° |
The shutter does not open, or is closed when:
-
The collimator is not in the right position.
-
The monochromator or X-ray port is not in the right position.
-
The beamstop is not mounted.
-
The shutter-open lamps are defect.
-
The radiation enclosure is not closed.
-
When using a rotating anode: the goniometer is not properly placed in front
of the X-ray port.
The X-ray high voltage is not switched on, or is switched off when:
-
The shutter does not close.
-
The high voltage cable is not connected to the X-ray tube (indicated by
a link, attached to the cable and connected to the tube holder).
-
The X-ray on lamp is defect.
Safety override
The system is supplied with two overrides, which are only allowed to
be used during technical maintenance (see safety
rules).
-
A key switch overrides the radiation enclosure safety. This state is signalled
by the flashing Open-beam lamp. When the lamp is defect the override is
cancelled.
-
Beamstop override vane. There are no extra warnings. See X-ray
safety.
Safety with rotating anode generator FR591
With a X-ray generator all safeties are handled by the generator, except
the shutter safeties: collimator and beam stop.
Safety signals
-
Shutter-open lamp: illuminated when the shutter is opened or is not fully
closed.
-
Shutter-closed lamp (only with sealed tube): illuminated when the shutter
is fully closed.
-
Open-beam lamp: flashes when the radiation enclosure safety is overridden.
-
X-ray lamp: illuminated when X-rays are generated.
With all lamps, excluded the green shutter-closed lamp, a defect lamp is
detected and activates the safety.
Safety with power off
Safety demands that circuits are automatically in a safe position when
they are out of order. So during power off, the goniometer Interface and
the X-ray high voltage are switched off by the safeties.
| Data-line to Host |
RS232: 300-19200 bd (auto-baud), 8 databits, 1 stopbit, no parity.
Galvanic separated. |
| Crystal illumination |
Supply 24 V, max. 80 mA, 2 W (not used). |
| X-ray / HV generator |
Hardware signals for shutter- and safety actions and a RS232 for software
communication. |
Environmental, according
to IEC 1010
| Environment |
Indoor use |
| Altitude |
Up to 2000 m |
| Temperature, operating |
20-22°C (within specification) |
| Relative humidity |
Maximum 80% if <31°C, decrease linearly to 50% at 40°C |
| Pollution degree |
2, according to IEC 664 (normal clean environment). |
| Protection |
IP 20, according to IEC 529 (limited protection to water). |
| Gasses |
No explosive gasses are allowed. |
| -- |
|
| Power supply |
Nominal 100V-120 V, 200V-240 V. max. 150 VA, single phase |
| Fluctuations |
±10% |
| Frequency |
47-63 Hz. 100V-110 V: 57-63 Hz. |
| Installation category |
II, according to IEC 664. |
|
|
| Goniometer |
146 kg |
| Interface |
8.5 kg |
Installation
Position
-
The equipment must be installed in a normal, clean and temperature regulated
environment and not in direct sunlight.
Power
-
The power must be connected by a detachable supply cord which is approved
by national authorities.
-
The power voltage is factory installed (see identification label on the
Interface) and can only be changed by a service engineer.
Interface
The Interface is mounted in the cabinet or placed on the table top.
The Interface must not be placed within 50 cm of the X-ray tube, to prevent
electrical problems due to spilt cooling water. Ventilation holes must
not be covered and no insulating or heavy object must be placed on the
equipment.
Safety ground
-
The equipment must be connected to the installation safety ground, e.g.
through a appropriate power cable.
-
The next safety ground must be checked: from the HV-generator to the X-ray
tube.
Connections and
controls
Operation
Operation of the MACH3 is executed through the computer program: see the
software manual. A limited number of operations can be executed by the
pocket- terminal.
Start-up procedure
-
Start up the computer program and wait till it notice: 'Connection is made
to CAD4 process'.
-
Switch on the power of the Interface (or reset with the button on the front
of the interface). The goniometer will slowly search for the zero reference
with an anti-collision routine.
-
The pocket terminal shows: 'Connecting to host computer'
-
After connection the pocket terminal shows shortly all axis positions zero
and then 'Zeroing procedure ongoing'
-
After finding the zero reference the Goniometer will position on zero.
The pocket terminal will give the position off all axis and the program
proceeds. The system is ready for use.
Pocket terminal
 |
F1 - F4 |
Function or submenu |
| ^, v |
Scroll through a setting |
| * |
Specials |
| Shutter |
Only active in Detector menu |
| <, > |
Position axis |
| STOP |
Stops all motors and closes the shutter. |
| Axes control |
Positioning of the four axes. Position in degrees |
| View positions |
Gives a choice of crystal align positions |
| Detector |
Displays the contrite of the detector. Open/close shutter. It is used
for technical alignment of the goniometer |
| - * - |
Special functions for maintenance |
| Other messages |
|
| Axes control |
|
| The +###.### * |
Theta position |
| Phi +###.### * |
Phi position |
| Ome +###.### * |
Omega position |
| Kap +###.### * |
Kappa position |
character * denotes : blank = motor stop, R = motor Running
| Submenu Axes control |
|
| Theta +###.### |
Position with < and > |
| Speed # |
Speed with ^ and v, blank: variable, from slow to fast
| 9 |
2000 °/min |
6 |
200 °/min |
3 |
20 °/min |
| 8 |
1000 °/min |
5 |
100 °/min |
2 |
10 °/min |
| 7 |
500 °/min |
4 |
50 °/min |
1 |
5 °/min |
|
| View positions |
|
|
| View 0 |
Positions to align the crystal |
View 0, 1 : Position in front of the goniometer
View 2, 3 : Position at the side of the goniometer
The choice between 0, 1 or 2, 3 is made in the host program in the
goniometer constants |
| View 1 |
|
|
| Phi +90 |
Rotates Phi 90°. Phi also rotates with < and > |
|
| Zero |
All axes 0° |
|
| Detector |
|
| ###### cnt/s |
Intensity, measured by the detector |
| Smooth ## s |
Average time: choose with ^ and v, default 1 |
| Dial aper #.# mm |
Dial: shows slits or aperture. |
| Attenuator OFF |
Attenuator on/off with F4. |
| Specials : * |
|
| Omega +90 degr |
Rotate Omega +90° |
| Omega -90 degr |
Rotate Omega -90° |
| Release motors |
Power off motors |
| Park |
Position for quick zero-find after power off (will also be executed
automatic during exit of the program). |
| Other messages |
|
| Connecting to host computer |
Tries to connect to the host program during start-up |
| Zeroing procedure ongoing |
Goniometer is running slowly to zero to determine the zero reference
position during start-up. |
| < Theta > |
Collision. See chapter Trouble shooting. |
Maintenance
Cleaning
For cleaning use a damp cloth, moistened with water or isopropyl alcohol.
No cleaning agent may enter the goniometer.
Tools
-
Absorber: Zirconium (for Mo) or Nickel (for Cu) filter, placed on the collimator
to attenuate the intensity of the primary beam to a measurable and save
level. It is only allowed to used with low kV/mA settings. Specification:
-
Mo: 7 foils of 0.108 mm Zr, attenuation approx. 2400x
-
Cu: 9 foils of 0.02 mm Ni, attenuation approx. 1900x
-
Test crystal: ammonium bitartrate.
-
Dial gauge for alignment of the X-ray tube.
Safety test
Shutter safety, with the FR590 or FR591.
-
Switch off the X-ray high voltage
-
Open the shutter. Check both shutter-open lamps.
-
Check if the shutter will not open after activating the next unsafe conditions.
The shutter must remain closed after reestablishment of the safety (except
the shutter lamps):
-
Collimator loose.
-
Beamstop removed.
-
Radiation enclosure open.
-
Shutter-open lamps removed.
-
Radiation enclosure safety override with the key-switch, Beam-open lamp
removed and enclosure open.
-
Check the flashing Open-beam lamp when the radiation enclosure safety is
override by the key-switch.
X-ray high voltage safety, with the FR590
-
Switch on the high voltage to minimum: 20 kV 5
mA
-
Check the X-ray on lamps.
-
Check if the high voltage falls off when:
-
X-ray on lamps removed.
-
Link of the HV-connector removed.
-
Shutter is manually opened by hand: take care
not to come in the vicinity of the primary beam!
X-ray high voltage safety, with the FR591
-
Switch on the high voltage on minimum: 20 kV 5
mA.
-
Check if the high voltage falls off when:
-
Shutter is manually opened by hand: take care
not to come in the vicinity of the primary beam!
-
See safety tests in FR591 service manual.
X-ray optics alignment
with a sealed tube
The measurement is done in a less safe condition : without beamstop and
the enclosure safety override. Safe operation is possible when executed
by a qualified person and precautions have been taken (see safety
instructions)
-
Don't let other persons come in the neighbourhood of the system.
-
Don't leave the system unattended, before reestablishment of the safety.
-
Don't come in the primary beam, although for security reasons it is attenuated
to a weak and safe intensity.
Measurement set-up
-
Place the 1-ring collimator.
-
Place the absorber on the collimator (take care: Mo and Cu use different
absorbers).
-
Position on Zero, aperture on open hole, no beta filter.
-
Remove the beamstop and place the safety switch override. take
care, the primary beam is measured, be sure Theta is always 0º or
use a second beamstop (X-ray shield) behind the goniometer (the
PVC windows are not sufficient to stop the primary beam), see X-ray
safety instructions.
-
X-ray generator:
-
Mo: 30 kV, 10 mA
-
Cu: 20 kV, 10 mA
-
For measuring the intensity use the pocket terminal : detector.
Adjustment
-
Adjust Tube 2Theta to 0°.
-
Dismount the monochromator and mount the X-ray port (two Allen screws {2.5
mm} at the bottom, which can be reached by two holes {not the middle one}
under the black monochromator holder). . During
replacement the power of the X-ray generator must be off.
-
Adjust tube 2Theta = 0° , and take-off = 3.0° (for fine focus 0.4
x 8 mm, for normal 1 x 10 mm focus : 6.0°).
-
Adjust on approx. maximum intensity with tube vertical (Allen 4 mm) and
tube horizontal (Allen 3 mm) adjustment.
-
Position slit horizontal and adjust on maximum intensity with tube vertical
adjustment.
-
Position slit vertical and adjust on maximum intensity with tube horizontal
adjustment.
-
If needed repeat both and lock the tube vertical adjustment (Allen 5 mm).
-
Remove the X-ray port. Switch off the power of
the X-ray generator during replacement.
-
Mount the monochromator and switch on the power of the X-ray generator
-
Adjust Tube 2Theta (Mo: 12.2°, Cu: 26.6°) and the Monochromator
2Theta on the anode target material used.
-
Position on slit horizontal and adjust on maximum intensity, first with
monochromator 2Theta, then tube 2Theta and monochromator height.
-
Position on slit vertical and adjust on maximum intensity with tube horizontal
adjustment.
-
Repeat both several times. Lock the monochromator height and tube 2Theta.
-
Mount the test crystal (Ammonium bitartrate), place the beamstop and remove
the absorber, use the 2-ring collimator, set X-ray generator: 50 kV, 20
mA (Mo and Cu).
-
Input top reflexes, or search for reflections, if the matrix is unknown.
-
Run program Align and adjust till the output is <0.01 (mm). This adjustment
will lower the intensity somewhat :
| TubeHo : adjust with tube horizontal adjustment with the
dial gauge against the tube head : Turning clockwise changes TubeH in positive
direction. |
 |
-
Mondeg :
-
adjust with monochromator-Theta adjustment: Turning upwards changes Mondeg
in positive direction (Mondeg = TubeVe/4). If this adjustment decreases
the intensity too much then first:
-
adjust monochromator height: Turning clockwise changes Mondeg in negative
direction.
-
Deterr :
-
EXPRESS:
-
Position Theta = -deterr / 8 (approx.)
-
In Specials, Goniometer constants, Motor feedback, do: Theta abs. zero
-
PC DOS/Linux :
-
KP, n 0 0 0 (n = -detererr / 8 (approx.))
-
GONCON, CASPAR, P, EN
-
TA, 0, TZ, 0, Q, Q, S, Q
-
Repeat Align
-
Lock monochromator 2theta.
-
Check the intensity with the Effic measurement
Adjustment points
EFFIC: intensity of the primary beam
The effic measurement is a relative measurement of the intensity of
the primary beam, which measures the total efficiency of the X-ray: tube
and optics. When the value is too low it is mostly caused by the tube (degradation)
or the X-ray alignment. The measurement is done during installation of
the system and later measurements can be compared with this value. Take
into account that because of tolerance in collimator and absorber (material
thickness) the values cannot be accurately compared with other systems.
Measurement:
-
Measurement set up: see X-ray alignment
-
Measure the intensity of the open beam. With new sealed tubes, fine focus,
this value should be approx.:
|
Mo |
Cu |
| with monochromator |
4500 cps |
8000 cps |
without monochromator
(no beta filter) |
7000 cps |
14000 cps |
Reset
When the goniometer program halts, it can be reset by the button on
the front of the interface. A reset is also possible by switching off the
power but this will also switch off the X-ray high voltage. See also the
start-up sequence.
Detector HV adjustment
Because the detector amplifier tube will degrade in time the HV (which
controls the gain) should be adjusted regularly (every 1-2 year). The detector
has two HV settings:
-
Hardware course setting in the goniometer, which is factory adjusted. This
needs only be readjusted with a new detector or HV unit.
-
Software fine setting, which can be used to compensate changes in time.
The range is large enough to accommodate Cu and Mo without changing the
hardware setting. The default is approx. 50 for Mo or 150 for Cu.
-
Measure the primary beam without beamstop with a standard absorber and
20 kV (Cu) or 30 kV (Mo) 10 mA (see safety).
When you have changed the tube, we advise to measure in the first stage
of X-ray alignment, without monochromator, to prevent confusing with misalignment.
-
Disable the automatic gain adjust in : GONCON, CASPAR, M, FL
-
First output FL and note the default settings
-
Fill in till the question: do you want pulse shift correction on : N
-
Change HV in GONCON, CASPAR till the intensity with LL = 100, WD = 250
is half the intensity with LL = low (Mo: 40, Cu: 30).
-
Input the default values for LL and WD (see table below).
-
Enable PULSESHIFT in GONCON, CASPAR, M, FL.
-
The other detector parameters are fixed:
|
GAIN |
LL |
WD |
| Mo |
200 |
40 |
130 |
| Cu |
200 |
30 |
140 |
User replaceable parts
X-ray tube replacement
-
See FR590 manual, chapter 4.2.2: Draining.
-
The tube holder is designed for an A-version (short) tube. When using a
B version (long) tube the 13 mm adapter must be used.
-
If the removed tube is used again : first measure the Effic
for later reference.
-
With a different anode: readjust the detector HV.
-
Realign the X-ray optics.
-
Maintenance of the tube: See FR590 manual, chapter 5.3: X-ray tube.
Monochromator
-
First switch off the power of the X-ray generator.
-
The monochromator is fixed with screws from below the collimator stand.
-
First adjust the Tube-2Theta before mounting.
-
With the X-ray port: first adjust Tube-2Theta before mounting.
-
When not mounted in the right position the shutter will not open.
-
After changing: realign the X-ray optics.
When mounting the collimator, the key must match the slot. When not mounted
in the right position the shutter does not open.
After changing, check if it is in good alignment with the collimator.
Second microscope (option)
 |
-
Attach the microscope to the base of the goniometer. Make sure that it
is perfectly vertical.
-
Mount the test crystal and centres it, using the fixed microscope.
-
Adjust the position of the second microscope as to get the test crystal
in the centre of the reticule, with adjustments height and rotate. Lock
the adjustments.
-
The microscope must be removed from the stand before operating the goniometer.
The stand does not interfere with any axis movement and can be left to
the base of the goniometer.
|
 |
Standard polaroid cassette (option)
 |
-
Position the goniometer with the POLA command. The polaroid cassette may
only be used in this position.
-
Mount the Polaroid cassette on the Omega axis.
-
Loading the film :
-
Set the LOAD/PROCESS lever to LOAD.
-
Insert the film pack firmly with 'this side toward lens' away from the
crystal.
-
Pull out the film pack until it stops. Pull the EXPOSE knob out. The film
is ready to expose.
|
Axial polaroid cassette (option)
-
Position the goniometer with the POLA command. The polaroid cassette may
only be used in this position. When Theta is accidentally moved from this
position the cassette arm shifts back.
-
Mount the Polaroid cassette on the goniometer base, opposite to the X-ray
tube.
-
Loading the film : see standard polaroid cassette.
Replaceable parts
Fuses must only be replaced with the power cable disconnected.
| Interface, primary |
4 A T, 4 x 20 mm, On the outside of the Interface. |
| Goniometer |
|
|
|
1.6 A T, 4 x 20 mm, F2 on the internal connection board |
|
|
200 mA T, 4 x 20 mm, F1 on the internal connection board |
24 V, 50 mA, 1,2 W, BA9s. Long life type. part number 2603.335 (order in
lots of 10). It is not allowed to install a LED lamp, because a defect
lamp is not detected by the Safety interlocks.
O-ring seals
-
Between X-ray tube and tube holder: part number 2132.017.
-
Water hose connection: part number 2132.018.
Test crystal
Ammonium bitartrate: (the crystal is different from a KappaCCD system :
larger size and oriented).
-
For goniometer head ECON and Eucentric: part number 0516.351
-
For goniometer head XYZ: part number 0516.352
Change: programmable IC for safety
(KappaCCD/MACH3 sold until 2002)
When the safety doors open the shutter will close for safety. When at
that moment the shutter does not close completely the HV will not be switched
off as under all other circumstances will happen. To prevent this problem
an electronic component (IC) must be replaced. This is not difficult and
can be done by an electronic engineer. It is not needed for KappaCCD/CCD
on a rotation anode generator.
-
Order (for free) the component called INT2E, part number 0540.656.
-
Open the goniometer interface from the front.
-
Remove the board with the three large connectors. First make a note which
cable connects to which connector.
-
Replace the component with note INT2A with the new component called INT2E.
The component can be removed by pushing it with a small screwdriver
under it. Take into account the position (a small hollow at one end). Take
care that not one pin is bend an not connected
-
You can also send this board to your local service for replacement.
-
Remount the board.
Ceramic tubes
When using Siemens ceramic tubes with the Nonius HV connector an isolation
tube is needed. This must be placed as shown below:
Accessories box
Troubleshooting
Start-up
and communication problems
-
With PC-DOS program, endless message: POLL: Unable to drop message due
to data overrun
-
This message can occur, even during several minutes, but at the end it
should stop and you can proceed
-
If not: start all over again from power down (including the Goniometer
interface).
-
With SGI-Indy computer: It is know that the RS232 port of the Indy is sensitive
for electrical disturbance, like power problems.
-
When the computer stops at: "Connection to CAD4 ..." and the pocket terminal
says: "connecting to the host computer", try the other RS232 port 2:
-
Login as CAD4
-
Command: cd data0/config
-
Command: mv rs232_port.ttyd1 rs232_port.ttyd2
-
Connect the small round RS232 connector one connector lower.
-
If this helps, then other port should be repaired by SGI.
-
Be sure to use the right start-up sequence
-
With
-
When you feel confident to check the electronics: open the Interface box
from top.
-
On the back board all red lamps (LED's) must be on.
-
Measure the +5V on the test pins on the back board. It should not be lower
then 4.8V. If it is not::
-
Disconnect and connect the front connector on the power supply: the upper
one on the right.
-
If this does not help, probably the power supply is defect.
Collision and maximum
angle switch
The anti-collision software prevents the goniometer from colliding with
the collimator holders or the axes to run past their limits. When a collision
takes place because of unexpected circumstances, like a technical error,
it is sometimes possible to return:
-
The pocket terminal shows a special menu: choose the axis you want to position
with F1-4.
-
First press the overrule button on the goniometer (near the connectors).
-
Then press simultaneously the pocket terminal < or >. Take care not
to run the goniometer further into collision.
-
As soon as the collision switch is released, the goniometer starts to zero-find.
Zero finding
Does not find zero during start-up
Check the dummy plug on the goniometer connector 'Limit switch'.
Very frequent zero finding during measurements
-
Problems with +5V power supply. When you feel confident you could check
the electronics
-
Technical error in axis positioning. Please mention the program and operating
system you are using and Service will ask you to read out some internal
test data.
Software error messages
which
can be caused by technical reasons are:
| 'Crash switch detected' |
The goniometer has touched a collision switch on a collimator holder.
Try to run out of collision. Check why
the system came into collision, because the software should prevent this |
| 'Generator problem' |
The X-ray high voltage is off. The generator will show the cause. Possible
reasons in the goniometer are:
The shutter does not close (also indicated by the shutter lamp). When this
happens the shutter must be repaired.
The high voltage cable is not connected to the X-ray tube. This is indicated
by a link, attached to the cable and connected to the X-ray on lamp.
The X-ray on lamp is defect.
|
| 'Shutter problem' |
The shutter does not open. This can be caused by:
The radiation enclosure is not closed.
The collimator is not in the right position.
The monochromator or X-ray port is not in the right position.
The beamstop is not mounted.
The shutter-open lamp is defect.
When the radiation enclosure safety is override with the key-switch:
Beam open lamp (with radiation symbol) is defect.
|
X-ray intensity problem
-
Check the EFFIC measurement and compare
it with earlier measurements. Mostly a low intensity is caused by the X-ray
tube.
-
Check the alignment of the goniometer.
-
X-ray measured while the shutter is closed: This
is mostly an oscillation of the detector amplifier caused by bad
connection.
-
Check if the detector parameters in the GONCON / CASPAR file are about
the same as in the delivery table.
-
Switch off/on the Interface and start again.
-
Disconnect and reconnect the two connectors on the detector.
-
Disconnect and reconnect the black (detector) cable between goniometer
and Interface.
-
Open the Interface and remove and remount the lowest (detector) board
-
Rotate Theta and move the detector cable by hand: : if the oscillation
stops if can be a broken wire in the cable.
-
Disconnect the round (HV) connector on the detector: does oscillations
stops?
Shutter does not open
-
The red shutter lamp goes on (temporary) but the shutter does not open
(green shutter lamp remains on, so both lamps are on). This error is mostly
caused by a safety error. Check in which part is the error:
-
Shift the collimator backwards as far as possible with the right
pin position.
-
Override the enclosure safety with the key (flashing lamp). If the
shutter now opens, check if all the doors are closed.
-
If safety door switched does not work: put thin plastic under the doors,
so they get higher, closer to the (magnet) switches.
-
Place new lamps in the red shutter on lamp.
-
Below the collimator holder there are two connections with a yellow and
black wire (the collimator safety). Short circuit them and check
if the shutter opens. If so: inform the Service department. For
safety : never use this short circuit permanent.
Manufacturer
and Service
BRUKER AXS B.V., P.O. Box 811, 2600 AV Delft, The Netherlands, Tel:
31 15 2152400 Fax: 31 15 2152500.
Service: see our web site: www.bruker-axs.com