August 2008
Introduction
The FR 591 rotating anode X-ray generator consists of a high voltage
supply and an X-ray tube with a rotating anode. The hot filament in the
cathode cup emits electrons that are accelerated by the high voltage. The
bias voltage together with the shape of the focus cup focuses the electron
beam in a small focal spot on the anode. The impact of the electrons on
the anode generates X-rays of an energy/wave length depending on the type
of material of the anode (Cu : 8 keV / 1.55 Å, Mo : 17,44 keV / 0.71
Å). Only a very small portion of the energy of the electrons is converted
to X-rays, the rest of the energy is converted to heat. The rotation of
the anode spreads the heat over a large area. This makes it possible to
use higher loads than on tubes with stationary anodes. The anode rotates
in vacuum and is internally cooled with water. A magnetic fluid seal around
the anode shaft maintains the vacuum while rotating. Continuous pumping
by a turbo molecular pump backed by a pre-vacuum pump maintains the high
vacuum.
The generator consists of :
-
Tube unit, with :
-
X-ray tube. The proper position of the X-ray shielding around the cathode
and the presence of the HV connector is monitored.
-
Two shutters. Shutter 1 (left hand) is the default shutter. Shutter 3 is
on the other ("water bridge") side.
-
Anode motor, which drives the anode by a belt. The anode current is monitored.
-
Turbo molecular pump, behind and in line with the tube. The controller
electronics is located behind the connector panel on the front side of
the tube cubicle.
-
Vacuum gauge on top of the tube (red cap) which measures the vacuum.
-
Safety key switches to disable the shutters or override the radiation enclosure
safety.
-
Two electric valves - one on top of the tube, the other one on the connector
panel - enclose a volume of nitrogen gas in the hose between them. This
volume is released automatically into the X-ray tube at power down. Flushing
the X-ray tube with dry and clean nitrogen gas shortens pump down time.
-
Connector panel.
-
HV unit, with :
-
ELCA module, a box with the electronics. The temperature of the electronics
is monitored.
-
High Voltage tank with a high voltage transformer submerged in oil. The
tank casing is the same for the 9 kW and 18 kW version.
-
Power generator. This can be 1 module (9 kW version) on one side or 2 modules
(18 kW version) on both sides of the tank. The temperature of the heat
sinks are monitored.
-
Oil pump which pumps the oil through the HV tank and cathode for cooling
these parts. The heat is transferred to the primary cooling circuit by
a heat exchanger on top of the tank. The flow, temperature and oil level
inside the tank are monitored.
-
Auxiliary unit, with :
-
Water pump to pump cooling water through the anode. The heat of this secondary
cooling circuit is transferred to the primary cooling water by a heat exchanger.
The primary water flow is regulated by an adjustable thermostatic valve,
mounted in the heat exchanger. It stabilizes the temperature in the secondary
circuit ( avoids condensation) and limits the flow when the output water
of the exchanger becomes colder (when less power is used). It is only useful
to limit water when using tap water. There is a small water vessel as a
buffer. The flow, temperature and level inside the water vessel are monitored.
-
Vacuum pump to take care of a pre-vacuum for the turbo molecular pump.
It is placed beside the Auxiliary unit.
-
Table / cabinet.
-
The half cubicle of the Tube unit is extended by one or two table cubicles
for carrying the X-ray measuring equipment.
-
The radiation enclosure is a shield to prevent access to the radiation
and to shield the low scattered radiation generated with regular diffraction
experiments.
-
The emergency switches on the outside of the table switch off the power
of the system.
Safety
instructions
Alterations, other than mentioned in this manual, can obstruct the safety
precautions and are therefore not recommended. This includes adjustments
which are sealed. Parts may only be changed by identical parts.
X-ray
safety
The equipment can generate open X-rays, which
can be hazardous when safety rules are not followed:
-
The equipment must be operated under the responsibility of an X-ray safety
officer, in accordance with local regulations.
-
The equipment must be operated, serviced and maintained by qualified personnel
who are trained on the hazards of and precaution against X-ray exposure.
-
The equipment contains safety circuits which should not be overridden,
other than with supplied: the enclosure key. These overrides may only be
used for service and maintenance, i.e. X-ray alignment.
-
The enclosure key should be kept under the custody of the X-ray safety
officer, not with the instrument. This also applies to the X-ray- or shutter
safety keys of the X-ray generator in case the X-ray port is not used or
X-rays should not be used.
-
Safety circuits must be tested by service personnel or the safety officer
at least every 12 months, as described in this manual (chapter
Maintenance).
-
Beam stop block's must be mounted on output ports not in use. The blocks
are delivered with the generator.
-
When the equipment is delivered without an enclosure, the X-ray safety
must be guaranteed by the user, by placing the equipment in a safety room.
-
The radiation enclosure is a shield for only the weak scattered radiation,
generated with normal diffraction equipment. When using other type of radiation
or large samples the shielding capacity must be recalculated and additional
shielding might be required.
-
Also take into account the X-ray safety of equipment installed.
Beryllium
The X-ray tube contains windows of beryllium. In normal use these area's
cannot be reached, but during maintenance, care must be taken not to touch
the window. When by accident the beryllium is touched, wash with soap and
water.
Fumes and dust from beryllium metal can be hazardous if inhaled. During
use, corrosion of the beryllium window may occur, but it should not be
cut, machined or otherwise removed. Disposal must comply with the national
and local regulations governing beryllium.
Electrical safety
Access to the area under the table is only allowed by qualified technical
personnel. Power terminals can be reached there on the outside of all modules.
When a qualified technician wants to access these parts, first disconnect
all automatic fuses and disconnect the mains supply before accessing any
interior part of the generator.
Specifications
| X-ray |
|
| Voltage |
5 - 60 kV |
| Current |
9 kW version : 2 -150 mA
18 kW version : 2 - 300 mA |
| Stability |
Voltage and current : 0.01% |
| Power (kW) |
9 kW (1 modules) or 18 kW (2 modules)
| Focus |
Type |
Filament |
W |
Cu |
Mo |
Ag/Au |
Fe/Cr |
| 1.0 x 10 mm |
BF3 |
broad |
15.0 |
15.0 |
15.0 |
15.0 |
15.0 |
| 0.5 x 10 mm |
BF2 |
broad |
15.0 |
15.0 |
15.0 |
11.0 |
9.0 |
| 0.3 x 3 mm |
FF3 |
fine |
8.2 |
5.0 |
5.0 |
3.1 |
3.0 |
| 0.3 x 3 mm HBA |
FF3 |
fine |
|
6.0 |
|
|
|
| 0.2 x 2 mm |
FF2 |
fine |
4.5 |
3.0 |
3.0 |
1.7 |
1.8 |
| 0.1 x 1 mm |
FF1 |
fine |
1.6 |
1.2 |
1.2 |
0.6 |
0.6 |
| 0.1 x 1 mm HBA |
FF1 |
fine |
|
1.5 |
|
|
|
0.5 x 5 and 0,1 x 2 are in the list but are not configured, so cannot
be used
HBA = High Brilliance Anode |
|
|
| Focus |
Focus types : see Power |
| Stability focal spot position |
<10 microns peak to peak |
| Focal height |
522 m above working space, optional 300 mm |
| Window |
0.25 mm beryllium |
| Take off angle |
0° - 10° |
|
|
The shutter does not open, or is closed when:
-
The shutter-open lamps are defect.
-
The radiation enclosure is not closed.
-
The goniometer is not properly placed in front of the X-ray port.
-
An extra external safety switch is open.
The X-ray high voltage is not switched on, or is switched off when:
-
The shutter does not close.
-
The high voltage cable is not connected to the X-ray tube (indicated by
a link, attached to the cable).
-
The tube cover is not in place
-
The tube shield (around the cathode) is not in place (indicated by a link,
attached to the shield)
-
The X-ray on lamp is defect.
Safety override
The system is supplied with two overrides. Their use should be limited
to technical maintenance (see safety rules).
-
A key switch can override the radiation enclosure safety. This state is
indicated by the flashing Open-beam lamp. When the lamp is defect the override
does not work.
Safety signals
-
Shutter-open lamp: illuminated when the shutter is open or is not fully
closed.
-
Open-beam lamp: flashes when the radiation enclosure safety is overridden.
-
X-ray lamp: illuminated when X-rays are generated.
With all lamps, a defective lamp is detected and activates the safety circuit
is activated.
Environmental, according to IEC 1010
| Environment |
Indoor use |
| Altitude |
Up to 2000 m |
| Temperature, operating |
°C (within specification) |
| Relative humidity |
Maximum 80% if <31°C, decrease linearly to 50% at 40°C |
| Pollution degree |
2, according to IEC 664 (normal clean environment). |
| Protection |
IP 20, according to IEC 529 (limited protection to water). |
| Gasses |
No explosive gasses are allowed. |
| Heat generation to air |
9 kW : max. 1 kW; 18 kW : max. 2 kW |
| HV tank oil |
Shell Diala D. The oil does not contain PCB's. This oil is also
used in FR571, GX21 and FR58.. generators |
|
|
| Power supply |
Nominal 200V-240V (phase - neutral), 347V-415V (phase =- phase). |
| Fluctuations |
±10% |
| Frequency |
50 - 60 Hz. ±5% |
| Connection |
Three phase Y (WYE or star), five wire (3 line, 1 neutral and 1 ground).
No earth leakage relay / protectors can be used. |
| Mains fuses |
9 kW : 3 x 35 A; 18 kW : 3 x 63 A |
| Power (kVA) |
|
L1 |
L2 |
L3 |
Total |
| 9 kW version |
4.3 |
4.7 |
6.2 |
15 |
| 18 kW version |
7.8 |
8.2 |
9.7 |
26 |
|
| Installation category |
II, according to IEC 664. |
|
|
| Water cooling |
|
| Flow and temperature |
-
Normal Cu or Mo anode (until 1998):
-
min. 15 l/min (9 kW, input 15 ºC) or min. 25 l/min (18 kW, input 15
ºC).
-
New Cu High Power Anode:
-
-
Minimum temperature : +15 ºC ( to prevent condensation)
|
| Water |
-
Primary water : normal tap water quality.
-
Secondary water : demineralized water.
|
| Input pressure |
2 - 6 Bar |
|
|
| Nitrogen |
Advised for a clean vacuum system. Max. 10 l /year. Pressure: 0 - 0,3
bar |
|
|
| Weight |
967 kg |
Operation
Start up procedure
-
Switch on the power with the key switch Power (left switch) on the control
panel.
-
In a few seconds all electronics are started and after this the power up
LED (yellow) goes on.
-
The generator is now is status POWER UP which means
the electronics are on, but all motors and pumps are still off.
-
Press the stdby button (left yellow button) on the control panel. When
pressing the red X-ray button the generator also goes to standby and proceeds
automatically to GENERATE. The start up sequence is :
-
Water pump starts.
-
After 2 seconds the oil pump starts.
-
After 2 seconds the anode motor starts to 600 rpm at the anode (the motor
about half this speed).
-
The tube is flushed with nitrogen for a few seconds.
-
The turbo molecular pump and pre-vacuum pump start are started.
-
When the turbo pump is above 50% of his final speed the vacuum gauge is
started.
-
Waits till the vacuum is below 4E-6 mBar.
-
The generator is now in status STANDBY, which means
it is ready to generate X-ray.
-
Press the red X-ray button on the control panel.
-
The generator is now in status GENERATE,
-
Increase the anode speed corresponding to the set power.
-
Switch on the HV generator, mA and kV will slowly rise to the setpoints.
When they have reached the set points a HV stable signal is sent to the
goniometer, indicating that the X-rays are stable and that data collection
can be started.
Shut off procedure
-
Press the red X-ray button on the control panel.
-
The HV switches off immediately.
-
The generator is now is status STANDBY.
-
Anode speed goes to standby speed (600 rpm)
-
Press the pwrup button (left yellow button) on the control panel or switch
off the power key switch (then after status POWER UP proceeds automatic
to power off.
-
Waits 0 - 20 minutes, depending on the previous HV power, to cool down
the tube.
-
Switch off the turbo pump.
-
Switch of the vacuum gauge.
-
Waits 2 minutes till the speed of the turbo pump is low.
-
Switch off the pre-vacuum pump.
-
A small amount of nitrogen is enters the tube.
-
After several seconds the anode motor, oil pump and water pump stop.
-
The generator is now in status POWER UP
-
Switch off the power with the key-switch Power on the control panel.
-
All power is switched off.
Shut down
There are several ways to shut down the generator :
-
Switch off the power with the key-switch Power on the control panel. The
generator will now shutdown in a controlled way, which will take time.
-
Emergency button. This will switch off all power at once.
-
Error. Depending on the type of error the status can be demoted (Generate
to Standby) or the generator can be shut off in a controlled or abrupt
way.
kV/mA set
-
The kV and mA are set from the control panel. After a new entry the value
will slowly go to the new setpoint. Limits are chosen depending on focus
size and anode material but the user can also choose lower limits with
Setup
/ Tube
-
The bias voltage is set automatically by the program, depending on kV and
focus cup. It can be changed in the mean menu Operate
/ generator. But when changing kV or mA the bias returns to the default
value.
-
The anode speed is set automatically by the program, as a function of the
power. It can be changed to a higher value with Operate.
But when changing kV or mA the anode speed returns to the default value
Control panel
The standard way to operate the generator is through the control panel.
Note that: the display windows of the buttons show the function of the
button (= the next action), not the present status. So open in the window
means that it will open when the button is pressed.
-
Key switch power to switch on the power. During the Standby and
Generate status this switch is by-passed by the electronics to shut down
in a controlled way, which makes it impossible to switch off at once using
this switch. Immediate shut down is only possible with the emergency buttons
at the front and rear of the table. The power key makes it possible to
prevent unauthorized access to the system.
-
Key switch X-ray to enable the generation of X-rays (not to switch
on the X-rays). The absence of this key prevents unauthorized X-ray generation.
-
Yellow button left. With this button the system can be switched
to the Standby status or back to the Power up status.
-
Yellow button Hotkey. With this button the most used display windows
are opened : Operate • Generator and View • Primary status.
-
Shutter 1 and 2 to close the shutter. The shutter can only be opened
(and closed) by the goniometer. With another electronic strapping (no Nonius
equipment) the shutter can also be opened by the control panel, but no
external control is possible. The opening of shutters can be disabled by
safety key switches under the X-ray tube, to prevent unauthorized opening
of the shutter.
-
X-ray to switch on/off the High Voltage. When pressed in the Power
up status the system is promoted to the Standby status and proceeds automatically
to the X-ray on status.
-
Enter / clear for the display commands.
-
Commands in the display:
Use the Hotkey to get the most used windows.
| Operate |
|
| Setup |
Setup parameters |
| Diagnose |
Not implemented |
| Service |
For special service commands, only with instructions given by Service
department. |
| View |
Show parameters (not adjust) |
| (status) |
Shows the status or the intermediate status (from which to which status)
POWER UP : only power on
STANDBY : all equipment running and ready to generate X-ray
GENERATE : generates X-ray, HV is on. |
| Operate |
|
| Generator |
Choose kV, mA or adapt pre-calculated bias voltage. It also shows the
limits (can be changed in Setup • Tube • Limits) |
| Recrystallise |
Start the automatic recrystallise program (ignore "done" before start) |
| Conditioning |
Start-up an automatic power up to the kV/mA limits, slowly rising the
power while checking the vacuum and trips. |
| Anode speed |
Choose higher then default calculated anode speed. |
| Setup |
|
| Shutters |
GCU enabled (default) : goniometer can use the shutter. It is not possible
to open or close the shutter from the control panel
GCU disabled : Close the shutter from the control panel. With special
electronic strapping also : open shutter.
Shutter 1 (left), 3 (right) |
| Generator |
Power modules used: number 1 , number 2, or two (2 modules). Don't
change this |
| Tube |
Choose focus (focal spot) size, anode material and kV/mA (user) limits.
The absolute maximum (tube) limits are calculated from the focus size and
the number of power modules. Be sure you have the management over the X-ray
(Setup • Management) and the generator in status Power up. |
| Management |
X-ray and Shutters default : Panel . When an external program wants
the X-ray management it takes over automatically and : "No management"
or "RS232" will show.
Service default : No management. Only when asked by Service department:
enter, choose obtain, press and release both enter and clear at about the
same time (enter a little bit before clear). This can be difficult. |
| Preferences |
Choose type of date (European, American).
Units : British (mbar and °C, default), American (psi and °F),
ISO (torr and °C).
Contrast (of display) :18 - 22, typical. Be careful! when choosing
a different value the display can disappear and this adjustment must be
found in the blind. |
| View |
|
| Primary status |
Show kV, mA, bias, vacuum and anode speed |
| Setup parameters |
Show shutter and anode. RS232 parameters are not right
Shutter: "GCU...": see Setup ; "Key-switch": shutter
key-switch is off. |
| Operation parameters |
Show on/off of : water pump, oil pump, anode motor and vacuum pump
value: temperature of oil, primary input
water and secondary output water circuit |
| Service |
For special service commands, only when asked Service department. For
writing: first get management. |
| Read |
Read of memory. It will default give the data of four successive addresses
in hexadecimal format, starting with input address (which will be supplied
by Service department) |
| Modify |
Write to the memory. Be careful : wrong writing can put the system
in a dangerous status, resulting in damage. |
| IO |
Reading and writing of IO registers. Be careful : wrong writing can
put the system in a dangerous status, resulting in damage. |
| Show revisions |
Show board revisions, adapted at the board with resistors |
| Ventilate |
To enter nitrogen in the tube. Opens the nitrogen NC valve for approx.
30 seconds. This must be done several times before the cathode can be removed |
Maintenance
General
warning
The mechanical maintenance described is intended for customers who have
attended a maintenance course. It comprises only a part of the required
periodic
maintenance. The skill of an experienced service man is required for
items related to high power, delicate mechanics and vacuum. When problems
arise with the described items please first ask Service department
for advise before using brute force. Bruker AXS cannot give free technical
support for work outside what is described in this manual and is not responsible
for the work, done by the customers.
Periodic maintenance
For a long lifetime of the generator a yearly preventive maintenance
is strongly advised. Some moving parts have a limited life and can damage
expensive parts when not replaced in time. A maintenance by Service consists
of:
-
Replacement of the ferroseal, filament, oil of pre-vacuum pump, turbo pump
lubrication, carbon brushes, V-belt.
-
Polish and clean the focus cup, recoil shield, anode, water bridge, vacuum
gauge.
-
Optional: refurbishment of the anode, or replacement by a new or a refurbished
anode.
-
Check of safety devices.
-
Several other checks.
Some items, mentioned in this manual, can be done by the customer after
a maintenance course, but there are limitations:
-
The anode can only be polished, but sometimes grinding is needed. Wrong
grinding can cause pollution to be pressed in the anode material, causing
a bad vacuum by high de-gassing. Also the deposit inside the anode should
be removed, otherwise the cooling becomes less efficient. This results
in overheating of the anode, which is shown by a bad vacuum caused by high
gassing. This refurbishment can only be done in the factory.
-
The recoil shield can only be cleaned in situ, because of the risk of wrong
remounting, causing vacuum leaks is too high.
-
For parts yearly needed see the part numbers
Filament
replacement
Filaments are supplied in a set together with mounting rod screws, clamp
segments, Allen keys and cathode O rings.
General
-
When mounting a new filament:
-
Clean parts should be handled with lint free, powder free gloves.
-
Filaments should be handled with plastic tweezers only.
-
All parts around the focus should be mounted by hand, never using a tool
other than the locking key, and never using force. When removing cannot
be done by hand, put some acetone in the screw-thread. If it is still not
possible to remove or mount the focus cup assembly, return the parts to
the factory.
-
Cleaning the cathode is occasionally necessary and should only be done
by a service engineer. Ceramic parts should only be cleaned with a diamond
file. Metal parts should be cleaned, first using Scotchbrite 07447 or 07448,
and then they must be rinsed and brushed with acetone. It is important
that no residues are left behind, since any residues will pollute the vacuum.
Preparations
-
When a filament breaks, the control panel will show a message and the generator
will go into the STANDBY status.
-
Go to status POWER UP.
-
After high power wait at least thirty minutes to cool down.
-
Ventilate the tube with the control panel under command Service. Because
this command only opens the valve 15 s, this should be done several times
(approx. 10 times) till the vacuum is gone.
-
Switch off the generator completely.
-
Remove the HV connector. First loosen the small Allen screw.
-
Confirm that the filament is broken by measuring the resistance in the
HV connector with (long) measuring pins between the COMMON and FILAMENT
contacts. A correct filament has a low resistance.
-

Fine focus
-
Remove the cathode.
-
With the adjustable cathode (not needed for the new Microstar fixed cathode)
-
Take care not to change S
-
Measure D:
-

-

-
Loosen the four Screws A, without removing them.
-
Unscrew the height adjustment nut, while lifting off the entire focus cup
assembly. A rocking motion may be necessary to remove the assembly.
-
Remove two uppermost screws B, which hold the filament in its mounting
rods and remove the filament.
-
If a small screw is
stuck, a way to remove is: Place the Allen key. Then get grip on it with
a side cutter (not a pair of pincers) an turn it out.
-
From the filament set, place two new Screws B with a new Allen Key.
-
Place the new filament with the leads in the grooves. The spacing of the
filament wires has a tolerance (maximum +/- 0.6 mm) so sometimes one wire
must be pushed gently into the groove.
-
Place the filament height setting jig and, using plastic tweezers, gently
adjust the position of the filament such that the filament sits between
the marks and against the top of the jig. Gently fasten the two screws
B. This is the most important step for a good alignment! When fastening
screws B the filament should not change position because there will be
tension on the wire which resultsin position change, like rotate, when
it becomes hot, causing a filament to cup error. If it does change while
fastening the screws, check if the filament wires are mispositionend and
check if the grooves are clean.
-
Place the entire assembly over the filament on the cathode, taking care
that the cup does not touch the filament. Screw the assembly in place.
-
Check the right position of the filament:
-

-
If the focus cup needs to be re-aligned to the filament, loosen the locking
ring (with the locking ring key), align the focus cup and tighten the locking
ring again.
-
Place the shorting key in the cathode.
-
Carefully place the depth gauge on the focus cup, avoiding any damage to
the filament. Adjust the height of the filament with the height adjustment
nut till the LED in the depth gauge is just on.
-
Tighten the four screws A so that the filament appears nicely in the centre
of the cup. Do not use force on the screws, as this can deform the mounting
tube, giving problems in removing it from the ceramic disk.
-
Turn the height adjustment nut a little bit back, so it is not tightened
but loose.
-
Once the filament was at operating temperature, it can be a little bit
curled, which is normally no problem. This is sometimes caused by the filament
and sometimes by wrong tightening of the screws B as explained above: all
must be tightened in the last stage.
-

Broad focus
-

-
Unscrew (not entirely) the four screws C.
-
Remove the focus cup.
-
Loosen the locking ring with the special key.
-
Unscrew the mounting tube and the locking ring entirely.
-
Remove the two upper screws B, which hold the filament and remove the filament.
-
Place two new screws B with a new Allen key, both out of the filament set.
-
Place a new filament in the mounting rods with a plastic tweezers and fasten
the screws B.
-
Pull the spindle out of the filament, if present.
-
Mount the mounting tube and the focus cup. Set the height, so that the
filament is in the focus cup. This is done by rotating the mounting tube
while keeping the focus cup by hand in position over the filament.
-
Place the shorting key in the cathode.
-
Fine adjust the filament height by using the depth gauge, so the LED is
just on. Be careful not to damage the filament. If one side is not correct,
this should be adjusted by lifting one of the filament feet in its mounting
rod. The rods can be reached through the holes in the mounting tube.
-
Lock the mounting tube with the locking ring and check the height again.
Readjust if necessary.
-
Lock the focus cup with the four screws C, while keeping the cup pressed
down on the mounting tube.
-
After reaching the operating temperature the filament will always be a
little bit curled:
-

Cathode mounting
-
With the adjustable cathode:
-

-
Remeasure D and if needed: X, Y and AC from the table (if not noted during
last replacement):
| focus |
0.1x1,
Microstar |
0.2 x 2 |
1.0 x 10 |
0.5 x 10 |
0.3 x 3 |
| AC |
8.5 mm |
9.0 mm |
9.0 mm |
9.0 mm |
10.0 mm |
-
Keep a log of the distances for future replacements: Y will never change,
X will only change after an anode replacement.
-
If D is within +/- 0.2 mm of the old value: remount the cathode without
changing the setting of S
-
If D is outside +/- 0.2 mm: calculate the new distance S = (AC + D + Y)
- X
-
FR591: and set the adjustable gauge for this distance.
-
Microstar: adjust the threaded collars and adjust this distance with the
slide-rule (delivered with the generator) before the cathode is placed
against it. The distance S should be as good as possible, but at least
with an accuracy of 0.2 mm.
-
With the Microstar fixed cathode:
-
Mount the cathode with the groove on the color, marked on the filament
box.
-

-
Mount the cathode carefully.
-
For horizontal tube and point focus the oil outlet is on top and the filament
is about horizontal. Take care: If by accident
the filament is mounted vertical the focus will be very thin and a groove
will burn in the anode.
-
Microstar: the filament must be visually horizontal.
-
Remove the shorting key.
-
Refit the HV cable connector and reconnect the interlock plug.
-
Fasten the small Allen key of the HV connector.
Recrystallization
-
Start the generator and start recrystallization from the control panel.
It will take about 10-20 minutes for the procedure to complete.
There is no tool for a more optimal horizontal alignment of the
focus. The previously used pin hole with fluorescent screen is not allowed
anymore because of the X-ray background radiation. When you want to use
your own safe tools, for instance using film inside a radiation enclosure,
you should get permission from the local radiation safety officer. The
radiation enclosure is designed to suppress only the weak scattered radiation
and not the high intensity of the primary beam.
When there is a possibility to view the focal spot, the shape will be:
Water seal replacement
The water seal in the water bridge needs to be replaced at least every
year.
-
Switch off the generator completely.
-
Remove the anode grounding brush, which is situated on top of the water
bridge.
-
Unscrew the three screws and take the water bridge straight off (when is
has two screws, it is the new high power anode, see below). The tubes are
not removed.
-

-
Press down the ring, for instance with two screw drivers, and replace the
middle black ring (the seal) in its rubber holder (this needs a second
person). Take care the rubber lip is properly positioned around the new
water seal. A spare seal is delivered with the generator.
-
Remount the anode grounding brush.
-
When this does not solve the leakage, the other part of the seal (the face
seal on the anode axis) should be replaced by a servicemen.
New High Power Anode
-
Disassembly:
-
Switch off the generator completely.
-
Remove the anode grounding brush, which is situated at the bottom of the
water bridge.
-
Remove the water bridge supply section by unscrewing two screws. The tubes
are removed.
-

-
Unscrew the inner stem retaining nut with the counterface seal key. This
is a left-handed screw, so turn clockwise to unscrew.
-
Remove the water bridge return section by unscrewing two screws. Take into
account that it fits very tight, so should be removed exactly straight.
If not, then it can stick. When it sticks, then push back and pull exactly
straight, perhaps with some wiggling.
-
Remove the water bridge water seal section by unscrewing three screws.
-
The water seal can be taken out by pushing.
-
Assembly:
-
Remount the water seal section and fasten the screws loosely.
-
Place the centring cylinder in place of the seal in the water seal section
and fasten the screws. Align also by rotating the section till the sloping
part is parallel to the part underneath
-

-
Sparingly lubricate the outside 0-ring of the seal with vacuum grease (delivered
with the generator), but no grease should come on the flat sides.
-
Place the guidance cone and mount the seal by hand. Don't use tools.
-
Remount the return section. Align it with the section underneath, so the
bottom, top and sloping sides are exactly aligned.
-
Fit the inner stem retainer nut (left handed) with a torque of 8 Nm. The
torque wrench is delivered with the generator.
-
With a new anode check if the anode rotates smoothly. If not, then dismount
and place a shim on the anode shaft (between anode and ferroseal,
two are delivered with new anodes).
-
Remount the supply section. Align it with the section underneath, so the
bottom, top and sloping sides are exactly aligned.
-
Remount the connector to the temperature measuring device and remount the
anode grounding brush.
Shutter
replacement
-
Switch off the generator completely.
-
Remove all covers.
-
Remove the shutter with the three screws.
-
Disconnect the wires. The connector is in the half cubicle, under the X-ray
tube. The colours and connection numbers are indicated nearby on a label.
When removing the cable, attach a wire or string, with which the new shutter
cable can more easily be drawn through the tubes.
-
Mount in the reverse way.
-
After mounting, check the shutter present (safety) switches in the control
panel : View • Setup parameters. When it shows "not fitted", shift the
shutter a little bit before tightening the screw, till the message disappears.
Anode
drive belt replacement
The anode drive belt should be replaced every year.
-
Remove the water bridge as explained in the Water seal replacement.
-
Remove the old belt.
-
Clean the pulleys with acetone.
-
Mount a new belt.
-
Adjust the tension of the V-belt by shifting the anode motor in the two
slots. The tension is correct when the belt can be pushed down about one
centimetre when applying a force of 0,5 kg near the motor. At too low tension
the belt will vibrate. Take care: with high tension the bearing of the
motor and anode will wear out.
-
Remount the water bridge as explained in the Water seal replacement
Anode
replacement
for Normal Power Anode (NPA)
General
-
This replacement can only be done by skilled personal, so by a service
men or after a course
-
Clean parts which are in vacuum should only be handled with non-fluff,
non-powdered gloves.
Rotorplate remounting
-
Go to the status POWER UP.
-
After high power, wait at least thirty minutes to cool down.
-
Ventilate the tube with the control panel under command Service. Because
this action only opens the valve 15 s, this should be done several times
(approx. 10 times) till the vacuum is gone.
-
Switch off the generator completely.
-
Take the drive belt from the motor pulley.
-
Disconnect the water hoses from the water bridge and rotorplate.
-
Unscrew the three large screws fixing the rotorplate to the head, while
holding the assembly in position.
-
Withdraw the assembly : first move it slightly in the direction of the
motor, then pull it straight out. The anode should not get scratched on
the way out. Place the assembly with the flat side of the anode on a clean
flat surface.
Anode disassembly
-
Remove the water bridge with three screws and remove the V-belt.
-
Clamp the anode retaining fixture into a bench vice and put the rotorplate
assembly on it, while carefully guiding one of the anode's six allen screws
(of the inner ring of screws) over the pin.
-

-
Unscrew the counter face seal with the counter face seal key and a 12 mm
wrench. Place the assembly on a flat surface once the counter face seal
is loose.
-

-
Take off the pulley and clamp ring.
-
Pull out the anode. Some force may be required here to overcome the friction
of the O ring in the bearing sleeve. Never use tools. If you don't succeed,
contact the Service department.
Anode cleaning
-
Cleaning of the anode is has limited use (see Yearly maintenance). When
desired, it is done with lint free clothes with acetone. Extreme pollution
can be removed with Scotchbrite 07447 or 07448. A convenient way is to
polish the surface while it rotates in a lathe. Proper care should be taken
to avoid scratches or damage to the shaft while it is clamped. After polishing
the anode must be washed with acetone.
Anode remounting
-
Place the anode with its flat side on a horizontal flat surface.
-
Carefully put the rotorplate assembly over the shaft, without an O-ring
in the bearing sleeve, just to check if it fits. The anode should fit smoothly
without any friction. Do not use force if there is friction.
-
Fit a sparingly lubricated (with vacuum grease) new O-ring (part number
2132162) into the groove in the bearing sleeve. Put also a tiny bit of
vacuum grease on the anode shaft.
-
Place the rotorplate assembly over the anode shaft while making sure that
this shaft is absolutely straight in the bearing sleeve. There should be
no force required until the O-ring is reached. A gentle tap on top of the
rotorplate should be sufficient to overcome the friction, after which the
rotorplate can be lowered to its proper position.
-
Place the clamp ring.
-
Fit sparingly vacuum lubricated O-rings (part numbers 2132162 and 2132163)
in the pulley and bring the pulley into position on the anode shaft. Take
care not to damage the O-ring while doing this.
-
Place the anode on the anode retaining fixture.
-
Mount the counter face seal and tighten it with a torque of 18 Nm.
Rotorplate mounting
-
Clean all parts with Scotchbrite 07447 or 07448 and clean afterwards with
acetone. Leave sufficient time to dry.
-
Put a slightly lubricated new O-ring (part number 2132963) to the head.
-
Carefully insert the anode into the head.
-
Shift it a little in the direction of the cathode to the right position.
Check if the O-ring is still properly positioned.
-
Fasten the rotorplate with its three screws.
-
Refit the belt.
-
Check if the inner water injector tube of the anode is properly centralized
before proceeding with the water bridge. If it is slightly out of the centre
it may just be bend back.
-
Remount water bridge.
Anode
replacement for High Power Anode (HPA)
General
-
This replacement can only be done by skilled personal, so by a service
men or after a course.
-
Clean parts which are in vacuum should only be handled with non-fluff,
non-powdered gloves.
Rotorplate remounting
-
Go to the status POWER UP.
-
After high power, wait at least thirty minutes to cool down.
-
Ventilate the tube with the control panel under command Service. Because
this action only opens the valve 15 s, this should be done several times
(approx. 10 times) till the vacuum is gone.
-
Switch off the generator completely.
-
Remove the anode speed sensor holder at the motor pulley.
-
Take the drive belt from the motor pulley.
-
Disconnect the water hoses from the water bridge and rotorplate.
-
Unscrew the three large screws fixing the rotorplate to the head, while
holding the assembly in position.
-
Withdraw the assembly : first move it slightly in the direction of the
motor, then pull it straight out. The anode should not get scratched on
the way out. Place the assembly with the flat side of the anode on a clean
flat surface.
Anode disassembly
-
Place the anode in the anode retaining fixture.
-
Remove the water bridge as described in the Water
seal replacement
-

-
Unscrew the counter face seal with the counter face seal key and a 12 mm
torque wrench.
-

-
Take off the pulley.
-
Pull out the rotorplate. Some force may be required here to overcome the
friction of the O ring in the bearing sleeve. Never use tools. If you don't
succeed, contact the Service department.
Anode cleaning
-
Cleaning of the anode is has limited use (see Yearly maintenance). When
desired, it is done with lint free clothes with acetone. Extreme pollution
can be removed with Scotchbrite 07447 or 07448. A convenient way is to
polish the surface while it rotates in a lathe. Proper care should be taken
to avoid scratches or damage to the shaft while it is clamped. After polishing
the anode must be washed with acetone.
Anode remounting
-
Place the anode with its retaining fixture.
-
Carefully put the rotorplate assembly over the shaft, without an O-ring
in the bearing sleeve, just to check if it fits. The anode should fit smoothly
without any friction. Do not use force if there is friction.
-
Fit a sparingly lubricated (with vacuum grease) new O-ring (part number
2132162) into the groove in the bearing sleeve. Put also a tiny bit of
vacuum grease on the anode shaft.
-
Place the rotorplate assembly over the anode shaft while making sure that
this shaft is absolutely straight in the bearing sleeve. There should be
no force required until the O-ring is reached. A gentle tap on top of the
rotorplate should be sufficient to overcome the friction, after which the
rotorplate can be lowered to its proper position.
-
Fit sparingly vacuum lubricated O-rings (part numbers 2132162 and 2132163)
in the pulley and bring the pulley into position on the anode shaft. Take
care not to damage the O-ring while doing this.
-
Mount the counter face seal and tighten it with a torque of 18 Nm.
-
Remount the water bridge as decried in the Water
seal replacement
Rotorplate mounting
-
Clean all parts with Scotchbrite 07447 or 07448 and clean afterwards with
acetone. Leave sufficient time to dry.
-
Put a slightly lubricated new O-ring (part number 2132963) to the head.
-
Carefully insert the rotor plate assembly into the head.
-
Shift it a little in the direction of the cathode to the right position.
Check if the O-ring is still properly positioned.
-
Fasten the rotorplate with its three screws.
Anode speed sensor
-
Refit the belt at the anode motor.
-
Adjust the tension of the V-belt by shifting the anode motor in the two
slots. The tension is correct when the belt can be pushed down about one
centimetre when applying a force of 0,5 kg near the motor. At too low tension
the belt will vibrate. Take care: with high tension the bearing of the
motor and anode will wear out.
-
Mount the speed sensor holder with the align tool and fix with the two
screws.
-

-
Mount the speed sensor and its coil.
Ferro
seal replacement for Normal- or High Power Anode
-
Remove the rotorplate assembly and anode as described in Anode
replacement.
-
Remove the bearing plate and all shims from the rotorplate.
-

-
Place the puller in vertical position on a flat table and put the empty
cartridge and the rotorplate over the shaft. The empty cartridge is delivered
with a new system and should kept for returning the ferroseal.
-

-
Place the large bush and the pin
-
Rotate the puller (hold the rotor plate by hand). The cartridge will be
pushed out of the rotorplate into the container. At the end the pin should
be placed in a lower hole, to pull out the ferroseal completely.
-
Check the rotorplate for burrs or damage. Make sure that the vacuum side
is clean.
-
Put a tiny bit of vacuum grease on the inner surface of the rotorplate,
but not in the water channels.
-
Take the cover from the new cartridge. There are two types of cartridges:
-
With a protruding plug at the bottom. Push with this plug the ferroseal
out a bit.
-
Without the plug. During mounting, first place thin distance holders, like
coins under the sides of the cartridge (take care:
not under the ferroseal). Pull out the ferroseal a little bit.
Remove the coins and proceed.
-
The protruding part of the ferroseal is needed as a guide into the bore
of the rotorplate.
-
Place the rotor plate and the new cartridge reverse on the puller and mount
the small bush and pin.
-
If needed pull out the ferroseal a bit as described above and remove the
coins.
-
Pull the ferroseal completely into the rotorplate. At the end the pin should
be placed in a lower hole, to pull in the ferroseal completely.
-
Remove the rotorplate from the puller and check if the cartridge is pulled
in completely. There should be no space left between the bottom of the
rotorplate and the first seal body.
-
Place the rotorplate back on the puller. Never
place the rotorplate on the table, otherwise there is a risk of damaging
the ferroseal.
-
Measure the protrusion D of the cartridge from the face of the rotorplate,
using a vernier caliper and determine the number of shims required. If
D is less than 0.10 mm and not zero then no shims are necessary. If D is
more than 0.10 mm then one shim has to be placed. If D is more than 0.20
mm then two shims, and so on. Place the (clean) shims.
-
Normally 3-4 shims are used. When you need more, then the seal is not pulled
in fully. Check for any obstruction. With too many shims the seal has free
space, can move and the O-ring does not fit. This can cause leakage and
defect of the ferroseal.
-
Place the bearing plate and clamp ring. Centre the bearing plate on the
clamp ring and tighten the bearing plate with three screws.
-
Remount the anode and rotorplate assembly as described in Anode
replacement
-
A ferroseal should be run in at standby speed for one hour.
Recoil
shield cleaning
-
Cleaning of the recoil shield should be done once a year, combined with
the replacement of the ferroseal.
-
This maintenance can only be done by skilled personal, so by a service
men or after a course.
-
Remove the cathode as explained in Filament replacement.
-
Remove the rotor plate as explained in Anode
replacement.
-

-
Remove the cathode holder.
-
Remove O-ring A.
-
Place the puller with the its parts and place the pin.
-
Rotate the puller till the end. The complete set can now (carefully) be
shifted out of the tube head and the recoil shield removed from the puller.
-
Remove O-ring B and clean the recoil shield with Scotchbrite 07447 or 07448.
After polishing the anode must be washed with acetone.
-
Fit a sparingly lubricated (with vacuum grease) new O-ring B (part number
2132529). This must be mounted from the widest side (not the output port
side) to prevent that parts, which are in vacuum are contaminated with
grease.
-
Remount the recoil shield by hand, so that the key in the tube head slides
into the groove of the recoil shield. Be careful, that O-ring B is not
damaged by the key.
-
When in good key position, push the recoil shield in by hand. Never use
tools.
-
Fit a sparingly vacuum lubricated new O-ring A (part number 2132634).
-
Mount the cathode holder.
-
Mount the rotorplate as explained in Anode
replacement.
-
Mount the cathode as explained in Filament replacement.
Anode motor
brushes
The four brushes should be replaced every year or when they are less
than eight millimetres long. (when new they are 23 mm long)
-
Remove the covers, first the top cover with the lamps.
-
At the back of the anode motor are four brush holders. Replace the brushes
one at a time (when loosening two, the wire connection is loose).
-
For a technician it is possible to open the motor:
-
Both the unit with the brushes and the plate on the side of the motor can
be opened to clean the motor.
-
First make a mark from brushes unit to motor unit to be sure to remount
in the same position.
-
Check if the brushes slide easy in their holders or clean it.
Vacuum
measuring probe for FR591 (red type)
The clamp under the vacuum measuring probe can be opened to enter air
into the head.
The vacuum measuring probe can be inspected when it does not start up.
Do this only when asked by the Service department. One can remove the measuring
probe by rotating it anti clockwise and lift (bayonet) without loosening
the clamp.
Checks to get the vacuum measuring head starting:
-
When the vacuum is too high:
-
Fill the head with nitrogen or with air by dismounting the probe with the
clamp and start again.
-
Touch the probe with your fingers. This will add enough contamination to
work for at least several months.
-
Check if the connector is all right.
-
Check if the probe is polluted. If so, clean it.
-
Check is the gap is approx. 0.4 - 0.7 mm. Be careful with the thin metal
foil.
When remounting, check if the probe is centred properly.
Pre-vacuum
pump oil change for FR591 pre-vacuum-oil pump
-
The oil should be replaced every year
-

-
Open the Out (side, bottom) and In (top) connection so the oil will run
out. The oil pump must be placed on a higher level.
-
Close the Out connection and fill the pump with 1 litre of oil. In the
indicator the level will be half way between min and max.
-
Close the In connection.
Shutter replacement
or exchange with shutter 3
-
Replacement
-
Be sure the power is off. The shutter can be removed with three screws
(2 long, 1 short). The shutter cable is connected to the connector rails
under the tube.
-
When mounting, first place the thin lead plate, then mount them together
on the tube. Mount with the screws loosely fitted.
-
Power on and read the shutter fitted switch (safety) in the control panel
: View • Setup parameters. Rotate the shutter till the status is not absent.
Then fasten the screws.
-
There is a list of cable connections close to the connector.
-
Exchange with shutter 3. When shutter 1 is defect it is possible
to exchange with shutter 3, which is mostly not used. This is not useful
when the shutter fitted switch is the error (shutter "absent").
-
Exchange shutter 3 and shutter 1.
-
Exchange the cable connections of shutter 3 and 1.
-
Don't exchange the GCU connector: this remains on shutter 1.
Safety test
Shutter safety, with the FR590 or FR591.
-
Switch off the X-ray high voltage
-
Open the shutter. Check both shutter-open lamps.
-
Check if the shutter will not open after activating the next unsafe conditions
:
-
Radiation enclosure open (and Open beam key switch disabled).
-
Shutter-open lamps removed.
-
Radiation enclosure safety (Open beam) override with the key-switch under
the table, Beam-open lamp removed and enclosure open.
-
Check the flashing Open-beam lamp when the radiation enclosure safety (Open
beam) is override by the key-switch.
X-ray high voltage safety
-
Switch on the high voltage to minimum: 20 kV 5
mA
-
Check the X-ray on lamps.
-
Check if the high voltage falls off when:
-
The X-ray on lamps are removed.
-
The safety connector of the HV connector is removed.
-
The safety connector of the tube shield is removed. After placing the connector
go to menu Setup • Tube and enter the same anode material again.
-
The main shutter is manually opened with a small screw in the centre of
the shutter indication: take care not to come
in the vicinity of the primary beam!
Technical
data
FR591
Microstar (Ultra)
FR591
| Position of (anode) water flow switch
in Microstar, FR591 new , FR591 old |
 |
FR591
Part number
list
Next items are advised to keep on stock or buy before yearly maintenance.
The table gives the numbers delivered with a new system in the spare parts
kit and the yearly use. Check if the part is for the normal power anode
(NPA) or the new high power anode (HPA). For expensive components it is
advised to buy when needed, to prevent the part is out of warranty when
it is used.
The next items we advise to keep on stock:
| Kit |
Use
yr |
Part number |
Description |
|
5 |
0571842
0571843 |
Fine focus filament set (5x) including mounting
parts and tweezers or
Broad focus filament set (5x) |
2
1 |
1 |
2985013
2134044 |
Water seal in water bridge, the standard replaced
item for NPA or:
Water seal for HPA |
| 1 |
1 |
3590003 |
Anode motor brushes, set of 4 |
| 2 |
1 |
2692023 |
Brush for anode grounding in the waterbridge |
| 5 |
1 |
4010071 |
Silicone oil and washer for the HV connectors |
| 5 |
|
2603050 |
Shutter lamp, inside the shutter. 5 V 60 mA |
| 10 |
5 |
2603335 |
Shutter lamp, on the head. 24V 50 mA |
| 10 |
5 |
2603212 |
X-ray on lamp. 12V 120 mA |
| 5 |
|
2603235 |
Open beam lamp. 12V |
| 1 |
|
4000014 |
High vacuum grease for O-rings |
|
|
0591114R |
Shutter, revised. Optional. This is an expensive
part, but take into account that because of the high radiation environment
and with high use the shutter is vulnerable to wear and for safety reasons
it is not possible to use a generator with one shutter missing, also on
a port which is normally not used. |
For customers who execute the yearly maintenance
(partly) themselves, we advise the next items will be bought:
| Kit |
Use
yr |
Part number |
Description |
|
1 |
Depends
on type |
Anode. It is possible to
-
Send the anode to the factory for refurbish (grind
when needed, polish, internal chemical cleaning)
-
Buy an anode and return the old anode.
-
Buy a refurbished anode and return the old anode.
This is only possible when on stock.
The lifetime of an anode is five years. |
|
1 |
0571355 |
Ferro seal. |
| 1 |
1 |
0571319
0168374 |
Counter face water seal, on the anode axis for
NPA or:
Counter face seal for HPA |
| 1 |
1 |
2033114 |
V-belt |
| 1 |
1 |
4010021 |
Oil for the pre-vacuum pump, 1 ltr |
| 5 |
2 |
2132162 |
O-ring 17.10 x 1.60 mm, for the anode pulley
and shaft |
| 5 |
1 |
2132163 |
O-ring 19.10 x 1.60 mm, for the pulley face seal |
| 5 |
|
2132310 |
O-ring 31.42 x 2.62 mm, for the water bridge
for HPA |
| 5 |
|
2132429 |
O-ring 37.77 x 2.62 mm, for the water bridge
for HPA |
| 2 |
|
2132528 |
O-ring 56.82 x 2.62 mm, between measuring flange
and head |
| 5 |
|
2132431 |
O-ring 47.0 x 5.34 mm, for cathode |
| 2 |
1 |
2132529 |
O-ring 58.42 2.62, for the recoil shield |
| 2 |
1 |
2132634 |
O-ring 66.35 2.62, for the recoil shield front |
| 5 |
1 |
2132963 |
O-ring 94.97x1.78, between the rotor plate and
head |
|
1 |
2140073 |
Grease cartridge for the turbo pump |
|
|
|
|
Technical warnings
Disposal
-
The HV tank contains PCB-free oil. A statement in Acrobat/.pdf form can
be retrieved
-
The Tube Controller Board contains a battery which should be regarded as
chemical waist according local regulations.
-
Beryllium windows
Troubleshooting
Startup errors
-
Generator does not go to the Power up status
(but no error is shown).
-
Display does not light
-
Check the red safety push buttons on the side of the cabinet (below the
table). They can be pressed by accident. Release them with the keys.
-
Check the manual fuse (elca) on the electronic
cabinet.
-
Check the main power (installation fuses or switches)
-
Open the electronic cabinet and inspect the two power supplies on the left.
Both should have a green LED burning. If not: then this power supply is
defect.
-
Display works, but Power on LED (yellow) is not on : Go to standby and
see what error message appears.
-
Generator does not go to the Standby status
(but no error is shown).
-
Check the status on the display :
-
"Power up" : check in menu Setup • Management X-ray should be set to "Panel".
-
"Standby - Power up" : the generator was going back to the Power up status
(for instance after an error) and will first finish this sequence before
going to the standby status again.
-
Check in menu View • Operating parameters of the Oil circuit is "ready".
If not : contact the Service department
-
Check if the X-ray on lamps are all right : replace them for new ones and
try again.
-
Shutter problem.
Low vacuum
-
A normal vacuum is:
-
1E-7 - 9E-7 without power.
-
below 9E-7 with medium power (for instance 3 kW)
-
below 5E-6 with high power.
-
A bad vacuum is mostly caused by leakage, which should be resolved by a
service man. Please check:
-
Check if the prevacuum pump is really running
-
Switch to standby (low anode speed and no HV). When vacuum is too low there
is probably a leaking O-ring. Because it is difficult to find which O-ring,
we advise a maintenance. In most cases it is the cathode O-ring. Take into
account that with every remounting of the cathode a new O-ring must be
fitted.
-
Adjust the anode speed to 5500 (will show 6000, takes about a minute to
speed up) but with HV off. When the vacuum is too low or not stable probably
the ferro seal is worn. Take into account that the ferro seal has a limited
life time of about a year.
-
Switch on the HV.
-
If a 13-16 (Vacuum alarm) occurs directly after HV on (before HV or mA
are raising) then check with the PC FR591 service program the error history,
if there are other errors preceding the vacuum alarm.
-
Set power on 20kV, 1 mA. When the vacuum is too low then probably the focus
cup is polluted.
-
Set power to maximum. When the vacuum is too low probably the anode is
polluted. This can be caused by:
-
The anode internal deposit is not removed for some time. This can only
be done in the factory during refurbishment.
-
Wrong procedure of cleaning the surface of the anode, giving pollution
pressed in the anode material.
-
Be sure you have not changed anything in the cooling water circuit, like
wrong connection of the cooling tubes.
Shutter does not open.
(sometimes error 13-08)
-
Shutter lamps defect or for Microstar lamp current is too low (service
must check)
-
For generators until 1999: if the HV shuts down the shutter is defect.
If the HV remains on it is a problem with safety or setup:
-
Check if the shutter lamps are on.
-
Look at the shutter status in the control panel : View • Setup parameters
-
"Absent" : The "shutter present" switch is not activated. Switch off the
X-rays. Ask an experienced mechanical engineer to loosen the shutter (screws)
a bit and rotate the shutter housing till the Absent sign disappears. If
this does not help the shutter has to be returned for repair.
-
"Key-switch" : The shutter key switch under the table is disabled. Switch
to position enable (key horizontal).
-
"Interlock" : One of the shutter safeties is active:
-
The door is not closed. Test by switching the Open beam key switch under
the table, so the Open beam lamp flashes. When the shutter can now be opened
the enclosure was not closed (detected by safety switches in the doors).
-
When using with Open beam : the Open beam lamp should flash. If not, replace
the lamp.
-
Check the Goniometer safeties : KappaCCD,
MACH3 or CAD4
-
Check if the goniometer detecting switch (hoover safety) is in correct
position (if present). This magnet/reed switch is mounted close to the
X-ray port /monochromator to detect that the goniometer is placed in front
of the X-ray port.
-
"Error" : An 13-08 error as occurred.
-
Check the shutter setup (also when generating X-rays) in the control panel
:
|
Shutter opened from goniometer |
Shutter opened from Control panel |
| Setup • Management |
shutters are set to "No management" or "RS232" |
shutters are set to "panel". |
| Setup • Shutter |
shutters are set to "GCU enabled" |
-
shutters are set to "GCU disabled"
-
When normally opened by the goniometer (setting on a board) and HV=on then
the shutter must also be opened by the goniometer (for the goniometer shutter
safeties)
|
-
Check if the X-ray port bayonet coupling (if present) is in good position.
The mechanical safety slot inside the shutter can prevent opening.
-
In the service program State bit = 1 and after HV on:
-
Sequence error:
-
Open the shutter by hand or:
-
Press the shutter open switch two times very quickly.
-
Now the State bit = 0 and the shutter will work till next powerdown/up.
This error can occur now and then is caused by timing error/bug
-
State bit = 1 directly after HV on
-
Check ST23/24 on the Tube controller board: When the new shutter PALs are
mounted (TUBE1CA, TUBE1CB, TUBE3CA and TUBE3CB, PCN24)
and when Nonius equipment is used then ST23/24 must be open.
Errors mentioned on the control
panel
Selection of errors from the control panel. For other error message
contact the Service.
Look at the second numbers of the error, for instance (0x13,
0x08) is 13-08 :
12-06 EMERGENCY PUSHBUTTON
The red emergency button on the outside of the cabinet is pressed
12-40 NOT STANDBY FOR HV
Not ready to generate HV (Generator is not in Standby status).
12-41 kV NOT RECHARGED
Action taken, for instance HV ON, when the kV wasn't fully discharged
after the HV OFF. Wait some seconds and repeat.
12-43 BAD VACUUM RECRYST
Starting recrystallization when the vacuum is too low. Wait till the
vacuum is high and the recrystallization will continue automatically.
12-44 AUTOMATIC FUSE AUX (Microstar)
The automatic fuse of the auxiliary unit (with pumps etc.) is off
13-01 LOOSE HT CABLE END
The "HT cable present" connector on the tube is not placed.
13-02 LOOSE TUBE COVERS
The white tube cover with the red lights is not in place, detected by
a reed switch with magnet in the cover. The signal is latched in a battery
powered circuit, so even if the covers are in place again or power has
been off, the electronics will show this error. The only way to reset is
to choose the (same) anode target (Cu or Mo) again from the control panel.
13-03 LOOSE TUBE SHIELD
The connector of the tube shield is not placed, indicating the metal
tube around the cathode of the X-ray tube is not in place. The signal is
latched in a battery powered circuit, so even if the connector is placed
again or power has been off, the electronics will show this error. The
only way to reset is to choose the (same) anode target (Cu or Mo) again
from the control panel.
13-07 TEMP ALARM COOLWATER
The temperature of the cooling water at the secondary output of the
heat exchanger is too high. Check the input water flow and the temperature.
The temperature is shown in menu View / Operating parameters.
After restoring the cooling the temperature of the secondary water can
still be too high, preventing the water pump to start. It has to cool down
on its own, which can take some time. It is possible to start the
water pump by a service command. Use this command only when needed, because
wrong use can start other parts without safety.
-
Get service management
-
Go to Service, IO and choose Write and Tube controller
-
Write (indifferent data) to address FF6B. The water pump will start.
-
Be careful: the pump is now running
without any safety, so after cooling down start again from power down.
13-08 SHUTTER FAIL
One of the shutters doesn't open, doesn't close or the safety shutter
fitted switch is not activated.
-
When the error is permanent one can track by reading the status in Service
menu, read, Tube-cntr :
-
Address FF5A, data: the last digit should be F ($...F). If not the
shutter present switch is not pressed. You could try to loosen the three
screws of the shutter a little bit and shift the shutter till it is F again.
If you don't succeed the shutter must ne readjusted in Delft
-
If the last digit = E then the error is in shutter 1
-
If the last digit = B then the error is in shutter 3
-
Address FF50 and FF54, data: the first digit should be 4 ($4..).
If it is E or 6 the shutter is not fully closed. You can close it with
the screw in the centre of the shutter rotation.
-
If it is address FF50 = then the error is in shutter 1
-
If it is address FF54 = then the error is in shutter 3
-
After restoring the error sometimes the generator must be powered down/up
to reset the error message.
13-0B ANODE DRIVE OVERRUN (FR591) or
ANODE DRIVE CURRENT (Microstar)
Start again and go in smaller steps to the kV/mA power.
13-0C VACUUM GAUGE FAILS
Vacuum gauge doesn't work. Possible causes:
-
Vacuum is too high. The gauge doesn't start when it is too clean.
-
Ventilate the head with nitrogen
and start again.
-
Check if you are not out of nitrogen.
-
Ventilating with air (loosen nitrogen tube).
-
With technical experience this can be done by disconnecting carefully the
vacuum
measuring probe with the clamp.
13-12 LEVEL ALARM WATER
Cooling water level in the vessel is too low. Refill the vessel with
demineralized water (first read electrical
safety)
13-13 FLOW ALARM TUBE
No water flow through the anode. Causes:
-
Water pump doesn't run.
-
Check if the water hoses for kinks.
-
During start after the water was filled : not enough water yet. Ask Service
for instructions to start the water pump and leave it running for at least
15 minutes, till the water flow doesn't switch off. Start again in the
normal way.
13-16 VACUUM ALARM
Check the low vacuum list
13-43, 56 TURBO SPEED FAILS or TURBO PUMP
TIMED OUT
Turbo pump speed too low or Turbo pump did not reach the right speed
after 4 minutes.
-
After several years of operating the Turbo pump bearings can be are worn.
A service man can check this by inspecting the grease cartridge . The advise
is to refurbish the turbo pomp, which is not expensive. One can choose
to do this through the local Balzers/Pfeiffer representative.
14-03 TEMP ALARM HEATSINK1
The heatsink is located in the power module, situated left of the HV
tank. This is cooled by two thin brown tubes.
-
Check if the primary water pressure is above 2 bar. A too low water pressure
is first noticed by this error.
-
Check the input and output water cooling connection on the power module.
Both must be cold. When the output is warm you can try to disconnect and
clean by flowing water through the power module in two directions.
14-0A LEVEL ALARM HV TANK
The oil level in the HV tank is too low. Check if there is oil leakage
on the floor.
14-0D XRAYS DISABLED
Trying to generate X-ray while the X-ray key switch is in position disabled.
14-0E XRAYS LAMP FAILS
Both red X-ray On lamps are defect.
14-42 FILAMENT SET FAILS
This can be caused by:
-
The tube current should not be too low: at least 10 mA.
-
Broken filament. When the generator is switched off and on it should come
with a broken filament message.
-
Many trips rapidly after each other.
-
Sudden manual change from bias high to low.
14-4B GATE DRIVE FAULT
-
For a 9 kW generator: Go to Setup, Generator and check if it is setup for
module 1. If not, change to that.
-
Check if the manual fuse 9kW1
(module 1) is on. If it switches off again there is a technical problem
and the Service department must be contacted.
-
The check circuit can be temporary switched off through the Control panel:
Service • Modify • Write: Gen-ctrl, RAM, address 1B7B, data 07. First get
service
management. Do this only after instructed by the Service department.
When this does not help, the electronics is defect.
14-4C FILAMENT BROKEN (FR591) or FILAMENT
CHECK FAILS (Microstar)
Filament current too low, most probably a broken filament
14-55 AUTO
The Stotz switch of a 9 kW module is switched off. Switch it on again.
14-56 SHORT
A short circuit between the filament and the bias cup. The filament
has to be replaced. If it is caused by a rotation of the filament, then
check with the next filament that it does not change during fastening (see
description filament replacement)
Internal
error: Input buffer overrun
This error is a communication error on the digital bus and is mostly
caused by electrical disturbance. This can be:
-
Strong HV trips. In time the generator head gets polluted and trips gets
more heavy, till they disturb the electronics. A maintenance visit is then
needed.
-
Power instabilities. After 1997 a new version of IC80 on the Main controller
board, called version MCN4C makes the FR591 less sensitive for very
short power dips. This helps however only when power dips are the cause.
Communication
problem with computer or goniometer
-
Check if the RS232 communication parameters are right with : Service •
Read • node : Main-ctl, address $81C6 and $81CA
-
For KappaCCD : (1200 bd, 8 data bits, 1 stop bit, no parity, no handshake)
-
$81C6 - $ 0 not used (hand shake)
-
$81C7 - $ 4 1200 bd, for 9600
bd: $ 25
-
$81C8 - $B0 1200 bd, for 9600 bd: $
80
-
$81C9 - $ 8 databits
-
$81CA - $ 1 stopbit
-
$81CB - $ 0 parity = none
-
If the values above are not setup (all different values) the software will
automatically choose 9600 bd.
-
Don't look to the view menu, because the RS232 parameters shown are not
right.
Anode motor brushes
wear
The anode brushes should last at least a year. When they wear in a too
short time you could clean the motor inside
and track the length for some time. When new, the brushes are 23
mm long. If cleaning does not help, probably the motor is worn and should
be replaced.
No mA
Open the tube and check if the filament does not touch the focus cup.
Service program
A service program is available which reads the internal registers and
values for trouble shooting the FR591. The program runs on a PC under Windows95/98.
Manufacturer
and Service
BRUKER AXS B.V., P.O. Box 811, 2600 AV Delft, The Netherlands, Tel:
31 15 2152400 Fax: 31 15 2152500.
Service: see our web site: www.bruker-axs.com