| KappaCCD |
 |
19 August 2008
part of 0540.712
Introduction
The KappaCCD consists of a X-ray source, three axes goniometer (Omega,
Kappa, Phi) to position the crystal and a Theta axis and DX to position
the detector. The two-dimensional detector measures the X-ray, which is
reflected in the crystal.
The X-ray can be generated by a fixed tube as part of the goniometer
and a HV generator or by a stand alone rotating anode generator. The X-ray
optics consists of a main shutter, a fast shutter, a monochromator to suppress
the ß- and brehms radiation and a collimator to get a small bundle.
Safety switches close the main shutter or switch off the high voltage in
unsafe conditions. In case of a rotating anode the main shutter and some
safeties are handled by the generator. A microscope displays the crystal
on the PC monitor. In the two-dimensional detector a phosphor converts
the X-ray to light, a fibre optic taper reduces the image and a cooled
scientific grade CCD measures the light signal.
The axes are decoded by precision optical encoders and are driven by
DC motors, which permit accurate, fast and smooth positioning with low
power. The system is controlled by a Server computer. A Host computer with
data collecting and evaluating software connects to the system. The goniometer
is surrounded by a radiation enclosure, which prevents access to the radiation
and suppresses scattered X-ray.
Safety
instructions
Alterations, other than mentioned in this manual, can obstruct the safety
precautions and are therefore not recommended. This includes adjustments
which are sealed. Parts may only be changed by identical parts.
X-ray
safety
The equipment generates open X-ray, which can
cause human hazards when safety rules are not followed:
-
The equipment must be operated under the responsibility of an X-ray safety
officer, as stated by local regulations.
-
The equipment must be operated, serviced and maintained by qualified personnel
who are trained with the hazards of and precaution against X-ray exposure.
-
The equipment contains safety circuits which should not be overridden,
other than supplied: the enclosure key and beamstop override vane. These
overrides may only be used for service and maintenance, i.e. X-ray alignment.
-
The enclosure key and beamstop override vane should not be kept with the
instrument but under the custody of the X-ray safety officer. This also
applies to the X-ray- or shutter safety keys of the X-ray generator in
case the X-ray port is not used or X-ray is not allowed to be used.
-
When using the beamstop override vane (only for alignment) the system must
not be left unattended. Theta always at 0°, safety door closed, low
kV/mA setting and a second lead beamstop.
-
The radiation enclosure is equipped with a second beamstop. It must not
be removed and the goniometer primary beam must be pointed to this beamstop.
-
Safety circuits must be tested by service personnel or the safety officer
at least every 12 months, as described in this manual (chapter Maintenance).
-
The collimator must always be installed.
Beryllium
The detector and X-ray tube contain windows of beryllium. In normal
use these area's cannot be reached, but during maintenance, care must be
taken not to touch the window. When by accident the beryllium is touched,
wash with soap and water.
Fumes and dust from beryllium metal can be hazardous if inhaled. During
use, corrosion of the beryllium window may occur, but it should not be
cut, machined or otherwise removed. Tube and detector disposal must comply
with the national and local regulations governing beryllium.
Specifications
| Angle, DX |
|
| Phi |
±180° |
| Kappa |
±175° |
| Omega |
±210° |
| Theta, with sealed tube |
-10°, +66° (in a 590 radiation enclosure) |
| Chi |
0° - 99.85° |
| DX |
25 - 170 mm |
|
|
| Speed |
|
| Phi |
0.0012 - 3000 °/min |
| Kappa, Omega, 2Theta |
0.0009 - 2000 °/min |
| DX |
2000 mm/min |
|
|
| Accuracy |
|
| Phi |
±0.015° |
| Kappa, Omega, 2Theta |
±0.010° |
| Reproducibility |
±0.005° |
| Resolution |
0.0003° |
| DX |
±0.5 mm |
| Reproducibility |
±0.1 mm |
| Sphere of confusion |
R = 5 µm |
| -- |
|
| Load, Phi |
max. 1 kg |
| Goniometer ECON head |
X, Y: ±1.5 mm, Z: 8 mm (For minimum DX: only ECON head and no
distance ring) |
| Microscope |
0.7 x 11 mm viewing area, grid approx. 0.05 mm |
Anti collision
-
Automatic anti-collision software.
-
Axis maximum position switches. Turning back is possible.
-
Collision switches on detector, collimator holder and beamstop.
X-ray optics
| X-ray source |
|
| Power |
maximum 60 kV, 300 mA |
| Crystal - focal point |
216.5 mm |
|
|
| Beta filters |
|
| Open hole |
beta/alpha = 13% |
| Cu |
0.020 mm nickel, beta/alpha = 0.2% |
| Mo |
0.108 mm zirconium, beta/alpha = 0.2% |
| Fe: |
0.016 mm manganese, beta/alpha = 0.2% |
| - |
|
| Monochromator, sealed tube |
Graphite, 12 x 6 mm, 0 0 2 reflection, mosaic spread approx. 0.4° |
| Collimators, sealed fine focus |
0.25 mm (351), 0.35 mm (352), 0.6 mm (353) |
| Beamstop |
1 mm, 1.5 mm, 2 mm |
|
|
| Screen |
63.5 x 63.5 mm |
| Pixels |
621 x 576 (binned) |
| Pixel size |
110 x 110 µm (binned) |
| Distortion |
<2 % |
| Phosphor |
25 mg/cm² Gd2O2S, optimized for Mo. Option optimized for Cu: 10
mg/cm² |
| Taper, magnification |
2.45:1 |
| CCD |
EEV, CCD05-30 grade 1, MPP (AIMO), 152 x 1242 (V x H) pixels |
| Dark current |
0.02 e/s.pix |
| Cooling CCD |
-60 °C |
| Cooling liquid |
typ. 0°C, min -10°C, Ethylene glycol / demineralized water,
50%/50% mixture |
| Read noise, total |
1 Mo photon at 200 kHz readout speed |
| ADC |
16 bit |
| Read time |
1.8 s (200 kHz, binned), 0.9 s (430 kHz, binned) |
| Pixel binning |
Default 2x2 in the CCD |
| Weight |
7 kg |
| Size detector |
20 x 15 x 15 cm (l x w x h) |
| Size Controller |
41 x 44 x 19 cm |
| Size Power unit |
35 x 14 x 22 cm |
| Cable detector - power unit |
4.6 m |
| Cable controller - computer |
7.6 m (maximum 46 m) |
| PC board |
ISA TAXI interface, DMA transfer |
Image defects (which are compensated for in software)
-
'Warm' pixels. A certain percentage of the pixels will have a higher dark
current than average.
-
Zingers. Spurious high intensity spots or stripes. These are mostly caused
by muons, generated by cosmic radiation, which traverse the silicon of
the CCD.
-
Distortion of the image. This is very low, caused by the taper.
-
Non uniform sensitivity, mostly caused by taper properties, such as separations
between fibre bundles (chicken wires).
-
A maximum of two black columns (grade 1).
Safety interlocks
The shutter does not open, or is closed when:
-
The collimator is not in the right position.
-
The monochromator or X-ray port is not in the right position.
-
The beamstop is not mounted.
-
The shutter open lamps are defect.
-
The radiation enclosure is not closed.
-
When using a rotating anode: the goniometer is not properly placed in front
of the X-ray port.
The X-ray high voltage is not switched on, or is switched off when:
-
The shutter does not close.
-
The high voltage cable is not connected to the X-ray tube (indicated by
a link, attached to the cable and connected to the tube holder).
-
The X-ray on lamp is defect.
Safety override
The system is supplied with two overrides, which are only allowed to
be used during technical maintenance (see safety
rules).
-
A key switch overrides the radiation enclosure safety. This state is signalled
by the flashing Open beam lamp. When the lamp is defect the override is
cancelled.
-
Beamstop override vane. There are no extra warnings. See X-ray
safety.
Safety with rotating anode FR591
With a X-ray generator all safeties are handled by the generator, except
the shutter safeties: collimator and beam stop.
Safety signals
-
Shutter open lamp: illuminated when the shutter is opened or is not fully
closed.
-
Shutter closed lamp (only with sealed tube): illuminated when the shutter
is fully closed.
-
Open beam lamp: flashes when the radiation enclosure safety is overridden.
-
X-ray lamp: illuminated when X-rays are generated.
With all lamps, excluded the green shutter-closed lamp, a defect lamp is
detected and activates the safety.
Safety with power off
Safety demands that circuits are automatically in a safe position when
they are out of order. So during power off, the goniometer Interface and
the X-ray high voltage are switched off by the safeties.
Environmental,
according to IEC 1010
| Environment |
Indoor use |
| Altitude |
Up to 2000 m |
| Temperature, operating |
20-22°C (within specification) |
| Relative humidity |
Maximum 80% if <31°C, decrease linearly to 50% at 40°C |
| Pollution degree |
2, according to IEC 664 (normal clean environment). |
| Protection |
IP 20, according to IEC 529 (limited protection to water). |
| Gasses |
No explosive gasses are allowed. |
| -- |
|
| Power supply |
Nominal 100V-120V, 200V-240V. max. 1500 VA single phase (included liquid
cooler) |
| Fluctuations |
±10% |
| Frequency |
47-63Hz. 100V-110V: 57-63Hz. |
| Installation category |
II, according to IEC 664. |
|
|
| Goniometer |
152 kg |
| Interface |
8.5 kg |
Installation
Position
-
The equipment must be installed in a normal, clean and temperature regulated
environment and not in direct sunlight.
-
The primary beam must be pointed to the second beamstop in the radiation
enclosure.
Power
-
The power voltage is factory installed (see identification label on the
Interface) and can only be changed by a service engineer.
-
UPS. For the CCD electronics and liquid cooler an uninterruptable power
supply is recommended to prevent the CCD temperature to rise fast in case
of a power failure. For a long lifetime a slow temperature gradient is
advised. The power connection is:
-

-
When the power is off, the CCD Controller is powered off, signalling the
power unit to slowly rise the temperature till it reaches the 0 ºC
of the liquid cooler. When the power returns before the UPS battery is
empty, the temperature will slowly return to -60 ºC. When power returns
after the UPS is empty, the liquid cooler will not restart automatically
and the CCD will not cool down.
Liquid cooler
-
The cooler should be filled with cooling liquid till half the indicating
glass (in the past it was custom to fill it till the maximum of the glass)
-
For a Julabo cooler, adjust to autostart after power up :
-
Press the key menu
-
Select with horizontal arrow : St
-
Adjust with vertical arrow : 0
-
Press enter and then menu key
Interface
Electronic interfaces are mounted in the cabinet or placed on the table
top. They must not be placed within 50 cm of the X-ray tube, to prevent
electrical problems due to spilt cooling water. Ventilation holes must
not be covered and no insulating or heavy object must be placed on the
equipment.
Safety ground
-
The equipment must be connected to the installation safety ground, e.g.
through a appropriate power cable.
-
The next safety ground must be checked: from the HV generator to the X-ray
tube.
Position of the Oxford Cryostream
A detector - crystal distance of 25 mm (using an ECON goniometer head)
is possible when a modified beamstop is used. Take care the cooling head
is not in the way of any goniometer movement.
 |
|
Positioning the system as shown in the figures. In this position the
angle between the vertical axis and the cooling head is about 10 degrees.
For DX = 25 mm a modified beamstop must be used.
Note that the dummy cooling head has a smaller diameter and is therefore
not well suited for sorting out the proper position. It is possible to
place a cuff around the dummy to make it suitable.
|
External Phi limit
switch for use with the FR559 crystal heating device
-
Preparing for installation . -----------------------------
-
You will need a (Windows 95/98) version of the KappaCCD server program,
dated after 26 March 1999 (go to the help in the main server window to
see this date).
-
You will need a interface software version dated after 26 March 1999. This
date is printed on an E-prom at the left side of the CPU printed circuit
board in the goniometer interface (there are two printed circuit boards
in the goniometer interface; the CPU board is the one with only one connector
at the front)
-
There are a series of settings that must be modified in the server program
set-up, some related to the phi limit switch unit, some to the special
goniometer head you will be using. It is therefore probably best to copy
your original settings file for re-use once you are finished using your
limit-switch unit. The set-up file is called KCCD.SET and is located in
C:\CCD\SERVER.
-
Mounting the limit switch unit.
-
Remove the black disk at the back of the Kappa block (there is a screw
at the centre of the block)
-
Remove the dripping tray at the base of the kappa support block.
-
Position phi to 0 degrees
-
Mount the switch unit using the support bracket and 2 screws. The switch
unit should be mounted at the right side of the phi axis :
-

-
Now the disks supplied must be mounted on the phi axis; the largest first
(phi should be at 0 degrees during this operation).
-
There is a cam in the smallest disk that fits a slit in the largest disk.
-
There is a Allen screw at the side of the smallest disk that can be loosened
during mounting.
-
It is important to position both disks correctly; the cam on the small
disk should be positioned close to the small vane that can be seen under
the switch unit (the vane has 2 micro-switches that sense the left and
right movement of the vane).
-
Because of the cam and slit this implies the protruding part of the biggest
disk will be positioned at the opposite side.
-
Once both disks are positioned and fixed to the phi axis the Allen screw
can be tightened.
-
In the server program go to setup->goniometer setup-> goniometer constants
and enable "phi limit switch unit".
-
Set the geometry of the goniometer head you will be using in setup->goniometer
head setup. For the 558 heating device set the top angle to 90 degrees
and the maximum radius to 30 (mm).
-
Shut down the server program.
-
Some additional settings in the C:\CCD\SERVER\KCCD.SET set-up file have
to be modified; these can be done using a regular text editor such as notepad
or edit (in case you are using a dos window). Look for the following parameters
and modify their contents as follows :
-
GON_PHI_MIN = -240
-
GON_PHI_MAX = 240
-
GON_DX_MIN = 45
-
GON_COLLISION_SECURITY = 3
-
PHI_MIN and PHI_MAX are phi axis limitations because of the limit switch
unit.
-
DX_MIN is the minimum distance that can be considered safe because of the
protruding parts of the heating device. If your goniometer head has a different
geometry, you might want to use a different value. If that case you can
start the server program using the standard value (in most cases GON_DX_MIN=25)
and position phi and DX (while keeping kappa = 0 ) until you find the proper
value. NOTE : the several motors of the goniometer are quite strong and
can do a lot of damage to the detector and/or goniometer alignment if a
crash occurs outside of the regular collision collars.
-
GON_COLLISION_SECURITY is the clearance in degrees the software keeps between
the kappa block and the detector/collimator/beamstop. The same remarks
as for DX_MIN apply here.
-
Fix the plug of the switch unit to the socket marked "Limit switch" at
the back of the goniometer. During standard use a dummy plug is used in
this socket; it must be re-applied after removing the limit switch unit.
-
Restart the server program.
-
Enough slack should be kept for the several cables running between the
limit switch unit and the phi axis device and the outside world, also when
moving omega and kappa. If the device tends to "strangle" itself it is
possible to limit the kappa range : modify in KCCD.SET the GON_KAPPA_MIN
and GON_KAPPA_MAX to appropriate values if necessary (default is -175 and
175 respectively).
Connections and controls
Red status led on the Power unit: (in the first
series the status led can be on another position)
| Off |
Cooling off or fault condition. |
| On |
Cooling switch on (after approx. 30 s) and cooling liquid at approx.
0°C or lower. |
| 1/4 on, 3/4 off |
Cooling switch off and CCD above approx. +10°C, ready to switch
off the power (or cooling liquid too warm to start cooling). |
| 3/4 on, 1/4 off |
Cooling switch off and cooling liquid below approx. +10°C: ready
to start cooling
or: Cooling switch on and cooling liquid above approx. +10°C: cooling
liquid too warm. |
Green led on the Power unit: power on of both Controller
and power unit. For test of only the power unit: check with the controller
powered off.
Operation
Start-up procedure
-
Switch on the Goniometer interface, Detector Power unit and controller
but with the cooling switch on front of the detector controller off. (When
the systems were already on: first switch off, then on)
Detector cooling on:
-
Switch on the liquid cooler to a temperature of 0°C.
-
After the cooler has reached this temperature, adjust the CCD temperature
on the Controller to -60°C (60 on the dial, not in the small window:
less than one turn). Never adjust lower than -60°C. (the red lamp on
front of the black Power unit should flash 3/4 on 1/4 of, when 1/4 on 3/4
off the liquid temperature is too high, when off the Power unit is in error).
-
Switch on the cooling switch on the Controller. (the red lamp on front
of the black Power unit goes on continuous, when off the Power unit is
in error).
-
Cooling down can take approx. 1.5 hr. The temperature is reached when the
green lamp near the temperature adjustment lights.
-
Start the Server program on the Server computer. The goniometer will first
look for the zero position (operation: GONINIT). When finished (operation:
none) the system is ready to get mastership from the Host computer (mastership:
none). It is also possible to work with the server program after setting
the mastership local with the command Acquisition · Grab mastership.
Shut off procedure
Detector cooling off:
-
Switch the cooling switch on the Controller off. After about an hour
adjust the liquid cooler to +20°C. The temperature is now slowly raised
by the power unit, to prevent a temperature shock. A fast warming up, for
instance during power failure will not damage the CCD, but for a long lifetime
a slow gradient is advised. Wait till the led on the Power unit switches
1/4 on, 3/4 off (after approx. 1.5 hour) before switching off the Controller
and Power unit.
Server software
The Host computer contains the programs to define the measurements, collect
the images and evaluate the data. It is connected to the Server computer
by an ethernet connection. The host computer can be a SGI under UNIX or
PC under Linux. During measurement the images are directly transported
from the Server computer to the Host computer.
The server computer is a dedicated PC for controlling the goniometer and
detector. After receiving the measurement parameters from the Host computer
it drives the goniometer and the detector. The Server computer runs under
DOS and Windows 3. The server program consist of:
-
Commands for technical service.
-
Input of goniometer and detector parameters.
-
Actions which are required near the goniometer, like aligning the crystal.
-
Possibility to run a measurement.
-
Bitmap utility to display measured images.
-
Program to display the image of the microscope.
-
Log window and file to save technical messages and diagnostic output.
It is not recommended to install other software, because it can disturb
the server program.
Calibration files and set-up file
The set-up parameters are saved in the Server computer in the set-up file
C:\CCD\SERVER\KCCD.SET. It is a text editable file, but most items are
changed through the Server program. Some parameters:
| ID_STRING |
Customer identification |
| DETECTOR_ID |
Last digits of the detector serial number |
| GONIOMETER_ID |
Last digits of the goniometer serial number |
| DET_PHOTGAIN |
Total photon gain in ADU/photon |
With a ; (semicolon) placed before GONIOMETER_PRESENT the server program
starts up without the goniometer. This can be used to test the detector.
The calibration files are saved in C:\CCD\CALIBR
| GRID.KCD (or .SPE) |
Image with holes on a 0.032" raster to calculate the image distortion
compensation. |
| NUNF...KCD |
Images from a Yttrium sample to calculate the non uniform sensitivity.
An image without X-ray is made (background) and two images with X-ray at
2Theta = +45° and -45°. |
The file C:\CCD\CALIBR\INFOLOG.TXT contains some calibration data, installation
data and other log remarks.
Upgrading Server software
Copy the complete upgrade diskette to the C disk. Overwrite existing files.
Don't copy read only files.
Maintenance
Cleaning
For cleaning use a damp cloth, moistened with water or isopropyl alcohol.
No cleaning agent may enter the goniometer.
Tools
-
Absorber: Zirconium (for Mo) or Nickel (for Cu) filter, placed on the collimator
to attenuate the intensity of the primary beam to a measurable and save
level. It is only allowed to used with low kV/mA settings. Specification:
-
Mo: 7 foils of 0.108 mm Zr, attenuation approx. 2400x
-
Cu: 9 foils of 0.02 mm Ni, attenuation approx. 1900x
-
Half-beam absorber: a 2 mm bar, placed in the crystal head.
-
Test crystal: ammonium bitartrate.
Safety test
Shutter safety, with the FR590 or FR591.
-
Switch off the X-ray high voltage
-
Open the shutter with Acquisition · Toggle shutter. Check both shutter-open
lamps.
-
Check if the shutter will not open after activating the next unsafe conditions.
The shutter must remain closed after reestablishment of the safety (except
the shutter lamps):
-
Collimator loose.
-
Beamstop removed.
-
Radiation enclosure open.
-
Shutter-open lamps removed.
-
Radiation enclosure safety override with the key-switch, Beam-open lamp
removed and enclosure open.
-
Check the flashing Open-beam lamp when the radiation enclosure safety is
override by the key-switch.
X-ray high voltage safety, with the FR590
-
Switch on the high voltage to minimum: 20 kV 5
mA
-
Check the X-ray on lamps.
-
Check if the high voltage falls off when:
-
X-ray on lamps removed.
-
Link of the HV connector removed.
-
Main shutter is manually opened by hand: take
care not to come in the vicinity of the primary beam!
X-ray high voltage safety, with the FR591
-
Switch on the high voltage on minimum: 20 kV 5
mA.
-
Check if the high voltage falls off when:
-
Shutter is manually opened by hand: take care
not to come in the vicinity of the primary beam!
-
See safety tests in FR591 service manual.
X-ray optics alignment
with a sealed tube
The measurement is done in a less safe condition : without beamstop and
the enclosure safety override. Safe operation is possible when executed
by a qualified person and precautions have been taken (see safety
instructions)
-
Don't let other persons come in the neighbourhood of the system.
-
Don't leave the system unattended, before reestablishment of the safety.
-
Don't come in the primary beam, although for security reasons it is attenuated
to a weak and safe intensity.
Measurement set-up
-
Zero all axes (this means Omega is 160°). DX = 150 mm. Always keep
Theta = 0°. No beta filter.
-
Place a #353 collimator.
-
Place the absorber on the collimator (take care: Mo and Cu use different
absorbers).
-
Remove the beamstop and place the red vane (safety switch override).
take
care, the primary beam is measured, be sure the second beamstop (X-ray
shield) in the radiation enclosure is in place
(the PVC windows
are not sufficient to stop the primary beam), see X-ray
safety instructions.
-
X-ray generator:
-
Mo: 30 kV, 10 mA
-
Cu: 20 kV, 10 mA
-
For measurements use in the server : Setup · Calibrate ·
Beam-alignment.
Adjustment
-
Adjust Tube 2Theta to 0°.
-
Dismount the monochromator and mount the X-ray port (two Allen screws {2.5
mm} at the bottom, which can be reached by two holes {not the middle one}
under the black monochromator holder). During
replacement the power of the X-ray generator must be off.
-
Adjust tube 2Theta = 0° and take-off = 3.0° (for fine focus 0.4
x 8 mm).
-
Adjust on approx. maximum intensity with tube vertical (Allen 4.0 mm) and
tube horizontal (Allen 3.0 mm) adjustment (measurement Continuous).
-
Raise the generator mA till the intensity is approx. 10000 ADU.
-
Mount the half-beam absorber (3 mm bar) on a goniometer head. Align it
with the microscope with / (former view 1)
and Phi 90° rotated: the edge of the bar must be in the centre (it
must absorb half of the beam).
-
Align horizontal and vertical :
-
Start a measurement (horizontal or vertical). Two measurements are executed
with Phi 180° rotated.
-
Both measurements must be equal within 5%. If not : adjust with tube horizontal
or vertical adjustment to the proposed "new" value (with measurement Continuous)
and measure again.
-
Lock the tube vertical adjustment (Allen 5.0 mm).
-
Test the alignment of the half-beam absorber :
-
Remove the half-beam absorber and measure Open.
-
Place the half-beam absorber and measure Vane. The error should be not
more than 25%. If it is : adjust to the advised new value and measure again.
When the alignment was bad, check the previous tube alignment again.
-
Remove the half-beam absorber and X-ray port. Switch
off the power of the X-ray generator during replacement.
-
Mount the monochromator and switch on the power of the X-ray generator.
-
Adjust Tube 2Theta (Mo: 12.2°, Cu: 26.6°) and the Monochromator
Theta on the anode target material used.
-
Adjust the intensity on maximum. First with monochromator Theta, then tube
2Theta and monochromator height.
-
Lock the monochromator height.
-
Measure the open intensity.
-
Mount the half-beam absorber and position horizontal with "Vertical" (don't
use the result of this measurement). Measure continuously and adjust with
monochromator theta till the intensity is half the previous measured open
intensity ±10% (a first coarse alignment).
-
As before, adjust the intensity within 5%:
-
Vertical : with monochromator theta (with measurement vertical)
-
Horizontal : with tube horizontal (with measurement horizontal).
-
Lock the monochromator Theta.
-
Set the X-ray generator to 10 mA and measure the open intensity (measurement
Open) without the half-beam absorber. This is the Effic
value.
Adjustment points
EFFIC: intensity
of the primary beam
The effic measurement is a relative measurement of the intensity of
the primary beam, which measures the total efficiency of the X-ray: tube,
optics and detector gain. When the value is too low it is mostly caused
by the tube (degradation) or the X-ray alignment. The measurement is done
during installation of the system and later measurements can be compared
with this value. Take into account that because of tolerance in collimator
and absorber (material thickness) the values cannot be accurately compared
with other systems.
Measurement:
-
Measurement set-up: see X-ray alignment
-
Measure the intensity op the open beam in the Server program with Setup·Calibrate·Beam-alignment
with measurement Open.
-
With new sealed tubes, fine focus, this value should be approx.:
|
Mo |
Cu |
| with monochromator |
3000 cps |
5500 cps |
without monochromator
(no beta filter) |
5000 cps |
9500 cps |
Revacuum
After some time the vacuum of the detector can become lower, resulting
in not reaching the low temperature (minimal after one year, mostly after
several years). The quality of the detector or measurements will not be
influenced by a lower vacuum. The vacuum can easily be restored with a
standard vacuum pump and the tube set delivered with the system:
-
Shut off the detector.
-
Connect a vacuum pump to the vacuum connection on top of the detector (an
oil filter is not necessary).
-
Switch on the vacuum pump till it reaches approx. 4 Pa (30 mTorr, 0.04
mbar).
-
Open the vacuum valve at the side of the detector (two full turns).
-
Wait till the pressure is at least the above value.
-
Close the vacuum valve with a torque of 3.5 - 4.0 Nm (approx. 30 Lbf-inches)
switch off the vacuum pump.
Reset
When the goniometer program halts, it can be reset by the button on
the front of the interface. A reset is also possible by switching off the
power but this will also switch off the X-ray high voltage.
Microscope alignment
Fine adjustment of the microscope is done
with two screws at the back plate of the holder. |
 |
Repair of
the small shutter
This repair can only be done by an mechanical engineer and only when
advised by the Service department
-
Part 1
-
Mount the steel ball, centre it and centre the microscope.
-
Remove the microscope including its square holder from the beamstop holder
with two screws (don't remove it from the holder).
-
Remove the black shutter cover (with the yellow led).
-
You will see now the shutter coil and plunger. Remove the plunger and clean
it with acetone. Clean the hole of the coil with a filt tip with
acetone.
-
Place the plunger and test with shutter open/close if it works all right.
When it does, go to part 3
-
Part 2. This part is done when a new shutter and alignment bar is send
from the Service department.
-
Remove the steel ball including the crystal head (is later used as a reference
for the microscope alignment)
-
Loosen Allen screw A and remove the shutter assembly.
-

-
Unscrew the coil and screw it on the new shutter.
-
Place the new shutter in such a way that the X-ray channel is in the direction
of the beam path and the locking screw (which must not be touched) is opposite
to the mounting screw A.
-
Place the alignment bar in place of the collimator with the small tip (which
fits in the shutter hole) to the shutter. Do not tighten the collimator
locking screw too much, so the alignment bar can slide.
-
Align the shutter assembly in such a way that the alignment bar fits inside
the X-ray channel and tighten Allen screw A.
-
The alignment tool must be returned to the Service department.
-
Part 3
-
Mount all items again.
-
Align the microscope on the steel ball.
-
The old shutter parts must be returned to the Service department.
Uninterruptable Power Supply
The batteries of the UPS have a life time of at least 3-4 years at 20º
C and a few power downs per year (at higher temperatures like 30 ºC
it can be half this time). After this time is it advised to switch off
the power (on the power rail inside the cabinet at the backside)
and check if the liquid cooler still runs after 1 hour and at least 2 indicating
LED's are on. Then switch on the system. If the time becomes too short,
the battery should be replaced.
User replaceable
parts
-
See FR590 manual, chapter 4.2.2: Draining.
-
The tube holder is designed for an A-version (short) tube. When using a
B version (long) tube the 13 mm adapter must be used.
-
If the removed tube is used again : first measure the Effic
for later reference.
-
Realign the X-ray optics.
-
Maintenance of the tube: See FR590 manual, chapter 5.3: X-ray tube.
Monochromator
-
First switch off the power of the X-ray generator.
-
The monochromator is fixed with screws from below the collimator stand.
-
First adjust the Tube 2Theta before mounting.
-
With the X-ray port: first adjust Tube 2Theta before mounting.
-
When not mounted in the right position the shutter will not open.
-
After changing: realign the X-ray optics.
When mounting the collimator, the key must match the slot. When not mounted
in the right position the shutter does not open.
After changing, check if it is in good alignment with the collimator.
Replaceable parts
Fuses must only be replaced with the power cable disconnected.
| Goniometer Interface, primary |
4 A T, 4 x 20 mm, On the outside of the Interface. |
| Goniometer |
|
| 24 V |
1.6 A T, 4 x 20 mm, F2 on the internal connection board |
| Crystal Ill. supply |
200 mA T, 4 x 20 mm, F1 on the internal connection board |
| Detector Controller |
|
| 100 - 120 V |
3 A T, ¼ x 1¼" (6.3 x 32 mm) |
| 220 - 240 V |
1.5 A T, ¼ x 1¼" (6.3 x 32 mm) |
| Detector Power unit |
|
| 100 - 240 V |
10 A T, ¼ x 1¼" (6.3 x 32 mm) |
Lamp
-
24 V, 50 mA, 1,2 W, BA9s. Long life type. part number 2603.335 (order in
lots of 10). It is not allowed to install a Led lamp, because a defect
lamp is not detected by the Safety interlocks.
O-ring seals
-
Between X-ray tube and tube holder: part number 2132.017.
-
Water hose connection: part number 2132.018.
Test crystal
Ammonium bitartrate: part number 0516.353
The crystal is different from a MACH3 system ( smaller size, randomly
oriented).
Troubleshooting
Start-up
and communication problems
-
PC-server program remains in: present operation = Goninit
-
Kappa does not move and Omega is slowly running to its maximum position:
check the dummy plug on the goniometer connector 'Limit switch'. Before
replacing first switch off/on the Interface.
-
Detector check failure 6,256, ADC
error or 'Specified AD Converter not installed'
-
Check the cable between PC board and controller, especially the connectors
-
Check if the PC board is good positioned in its connector. This can
be critical. Remount the board.
-
Add the line SKIP_ZERO at the top of start-up file C:\CCD\SERVER\KCCD.SET
with notepad. (take care to remove this
line after testing !)
-
Reset the interface. After start of the server program:
-
Program does not ask: "Do you want to skip zero?":
-
Check again with power off/on (the reset button does perhaps not work).
If still not asking then there is probably a RS232 communication problem.
Proceed:
-
Check +5V power supply When you feel confident
to check the electronics: open the Interface box from top and measure the
+5V on the test pins on the back board with the power indicating LED's.
It should not be lower than 4.8V. If it is not::
-
Disconnect and connect the front connector on the power supply: the upper
one on the right.
-
If this does not help, probably the power supply is defect.
-
Check the two LED's on the Interface Processor board. Both must be off,
the left one blinks. After power-up the processor starts to autobaud: sends
messages with different baudrate to the computer with shows as blinking
of the left led (Send, every 5 s). The computer answers (right led, Receive),
after which the communication is established (both fast blinking for a
short time).
-
Present operation = none (system has started) : It is a problem in the
goniometer.
-
Read the diagnostics in Service program: Goniometer, diagnostics. Check
"Omega limit switch reached" which should be 0. If it is 1 please consult
the Service department.
-
Update the position in the service program. Take care: it takes his present
position as zero and anti-collision software does not work.
-
Position every axis +/- 5 degree and check which axis does not position
right.
-
Check in SETUP • Goniometer • Diagnostic if there are Positioning or Index
errors and inform the service department.
-
No communication with the goniometer: PC-server program is not executing
GONZERO
but directly goes to NONE:
-
Check in server program: Setup, Goniometer, if the goniometer is present
(tagged). If not: edit the file \kccd\server\kccd.set and add a line GONIOMETER_PRESENT
(tagging in the setup does not help because of a software bug)
-
Host computer
-
Message: "Could not make connection:need mastership to cryostat".
-
Check the settings in the Server program: Setup, Network, Automasterset
to Network should be tagged.
-
Software bug. Use the newest server software upgrade
No acquisition possible / CCD not
present
If after grabbing mastership the acquisition image in de Server program,
remains grey:
-
Check in Setup if CCD is present.
-
If not, force CCD present and then read the CCD parameters: if the frequency
is blank (it should give 200 kHz) then the PC cannot connect to the detector
controller. Check:
-
Is the power of the ST138 on.
-
Check the thin cable between PC and detector controller. Remove and reconnect
the connectors.
-
Be sure the cable is connected to the TAXI board, because sometimes it
is by accident connected to a COM port.
-
Open the PC and remove/replace the TAXI board to be sure it is right positioned.
-
Check in Windows if the Taxi board is installed: My computer, Control panel,
System, Device manager, Data Acquisition, Princeton Instruments PCI Interface,
General.
-
Some eccentric errors:
Collision and maximum
angle switch
The anti-collision software prevents the goniometer from colliding with
the collimator holders or the axis to run past their limits. When a collision
takes place because of unexpected circumstances, like a technical error,
it is sometimes possible to return:
-
On the Server computer a menu appears.
-
Choose the axis and direction to run out of collision.
-
Press the overrule button at the back of the goniometer. The axis now positions
a small step.
-
If not enough: repeat. As soon as the collision switch is released, the
goniometer starts to relocate the zero position.
-
Continuously detector collision:
Remove the black collision ring (with an easy click) and dry and clean
it inside. This error especially occurs with using a cryostat, which can
cause condensation in the ring.
Check
the position
CCD temperature problem
-
First check the temperature of the cooling liquid (approx. 0 °C). If
this is not OK:
-
Check if there is enough ventilation.
-
Warm up the detector and switch off. If the liquid temperature is still
too high, the error is in the liquid
cooler.
-
Start up again.
-
Rotate the temperature adjustment on the CCD controller till the led is
green: this is the real CCD temperature.
-
If it is slightly higher then -60 then probably the detector need to be
revacuumed.
For the moment it is possible to work for instance at -55. Most detectors
has a very low dark current which, because of the automatic background
compensation, will not influence the quality of the measurement at higher
temperatures.
-
If the temperature is very high, the cooling could be switched off.
-
Switch off the ST138 and check if the green front led of the black power
unit is on. If not:
-
Check the fuse at the back of the power unit. If this is right of blows
again after replacement, then inform the service department
-
Switch the power unit off and on and check the red front led, which should
remain on. If it switches off, please check the error on the right side
of the Power unit. Through a slit three of four LED's are visible (not
with some of the first series). With four LED's normally the most left
one is on (is the same as the front red led) and the one next to it flashes.
In case of error the three right ones give an error code. Please note which
is on and off and consult our service department.
-

-
Leakage of
cooling liquid
-
Small leakage at the coolers site: Although the cooling bath is closed,
condensation is possible at high humidity (must higher than our ambient
specification). When the cooler was filled till its maximum, spill over
will occur. Be sure to fill the cooler in
the right way, so you can check when it is full. Take into account the
liquid should not be diluted too much with water, otherwise it will freeze.
-
High leakage can occur in older equipment, at the connection of the flexible
tubes (to detector or cooler) to the metal tube (in the radiation enclosure.
For fast (or temporary) repair, open it and try to reconnect. The Service
department can send you clamps for this. The connection of the detector
tubes are at the position where the tubes enters the roof.
Error in the image
-
Light area in the right part of the image when measured with high
Theta : this is X-ray background scattered from the beamstop (only several
ADU).
-
Image broken in blocks. The image is wrong written in PC memory.
-
Check the connectors of the cable between CCD controller and the PC.
-
Check if the PC board of the CCD controller is fitted in the PC in the
right way.
-
Run a performance measurement in Setup·CCDsetup·CCDtestsuite.
It is a test without X-rays.
-
The ADC-zero will be between 50 - 150 with 200 kHz. Other values have no
influence on the measurement, unless it is lower than 10.
-
The readout noise is typical less than 2.5 with 200 kHz.
-
The dark current is typical less than 10 with the 600 s measurement. The
dark current is so low, that even a higher value will have a minor influence
on the measurement quality.
-
Check a flood field measurement with the Yttrium sample.
-
Use a 353 collimator. Position Theta = +22.5° and Phi = +45°. The
sample should be parallel to the detector surface. Measure for at least
10 minutes. take care, the radiation is high in
all directions. It must be measured with a closed X-ray enclosure
-
Mo: Yttrium sample, 50 kV 40 mA.
-
Cu: Fe sample, 45 kV 32 mA.
Generator down (software message)
The X-ray high voltage is off. The generator will show the cause. Possible
reasons in the goniometer are:
-
The shutter does not close (also indicated by the shutter lamp). When this
happens the shutter must be repaired.
-
The high voltage cable is not connected to the X-ray tube. This is indicated
by a link, attached to the cable and connected to the X-ray on lamp.
-
The X-ray on lamp is defect.
X-ray intensity problem
-
No X-ray (but no error messages):
-
Check if there is also no X-ray with the Effic
measurement.
-
If the main shutter opens (audible and visible) and the led on the small
shutter is on, then probably this shutter should be cleaned
of replaced.
-
X-ray too low:
-
After mounting a new (Siemens) tube: check if the tube is right. It should
have a part number ending with -180 and not -90 (which gives a line focus).
-
Check the Effic measurement and compare
it with earlier measurements. Mostly a low intensity is caused by the X-ray
tube. When you want to be sure the error is not in the detector you can
check the intensity of the primary beam with a separate detector (for instance
a scintillator detector usable for 8-18 keV).
-
Check the realignment of the goniometer.
Perhaps the primary beam doesn't go through the centre of the goniometer.
It is also possible to test it by using a small 0.4 mm steel ball on the
crystal position and make an image of the primary beam (Always
with absorber and low HV/mA setting)
Crystal centring with
liquid nitrogen
To prevent changing of the crystal centre when using liquid nitrogen:
-
Use a metal crystal head.
-
Use the brass mounting pin
-
In collect: use the low temperature limitation. If not:
-
The glass capillary maximum 5 mm.
Problems with liquid
cooler or Uninterruptable Power Supply
If the equipment is out of warranty the repair can be carried out by
the local representative. The addresses can be found on the web:
Manufacturer
and Service
BRUKER AXS B.V., P.O. Box 811, 2600 AV Delft, The Netherlands, Tel:
31 15 2152400 Fax: 31 15 2152500.
Service: see our web site: www.bruker-AXS.com