Bruker AXS index previous next

Previous: collect: Data collection strategy and data collection
Next: Cell determination using Denzo


Data collection

Collect, data collection experiment

For a datacollection experiment, two buttons appear in the experiment window:

You can either run through these 2 steps in sequence, or just press the "collect data" button which will run through the whole process automatically. Each button will turn green when the associated action has been completed.

There is one advantage of doing it in 2 steps: it makes it possible to "Save" the experiment using the "File" menu before the actual data collection is started. Doing so will make it easier to recover from a power failure or a system crash.

In this example we will start from a directory where we have already performed a successful cell determination. As you will see later, it is possible to perform a data collection without first determining the unit cell orientation matrix. However, some questions might give other defaults and/or be asked at different moments.

We start by pressing the "Make Scan-set" button. A dialog box is shown asking you for information about the data to be collected.

Data Set parameters

Maximum diffraction angle theta
For a Mo-K-alpha dataset, a normal dataset is complete to 27.5 degrees theta, equivalent to a full sphere for Cu-K-alpha (~0.79 Angstrom resolution). If a higher resolution dataset is required (e.g. for electron density deformation studies), this value can be raised. If the crystal diffracts very badly, this value can be lowered. Results from the findresolution program will be used to fill in the default value; if that program has not been run, it is filled in from a configuration parameter.
Under "More options" another two questions appear:

Data set parameters; expert view

Minimum diffraction angle theta
Due to the beamstop, some very-low resolution reflections can sometimes be occluded. That is why this value defaults to 1.5 degrees and not 0.0. It can be increased if an efficient datacollection is desired to collect a shell around a previous Ewald sphere (e.g. you already have a data collection up to 27.5 degrees, and now want to collect the rest upto very high resolution).
Reduce cell axes to not lower than
For large unit cells, a large data set will result. Simulating all these reflections will take a lot of memory and calculation time. Using this parameter, collect will reduce the cell dimensions by integer factors (hence leaving out intermediate reflections) for large cells such that the data set size is reduced, and strategy calculations become faster. 10 Angstrom is normally sufficiently small on one side, and sufficiently accurate on the other (For radiations other than Mo, appropriate scaling will be performed on the default).
Both the maximum and minimum diffraction angle can be requested in degrees (default) or in "Angstrom resolution". This is determined by a configuration parameter.

Once you have confirmed the settings using "OK", another dialog will pop up asking you where to get the unit cell parameters.

Unit cell parameters

Two paths can now be followed:

  1. You know the unit cell and orientation, and want to use it for the strategy.
  2. You simply want to measure a half sphere or a full sphere.
The second case is easy: press the "Triclinic" button.

If you want to make use of the symmetry and orientation of the unit cell, you will need:

See the section about the "index" experiment for a way to obtain a suitable pair of ".x" and ".kcd" files, or the section about the "phi/chi" experiment for a way to obtain a ".rmat" file.

Under "More options", you will find two more parameters:

Unit cell parameters; expert view

Point group of data set
If this is set to "Default", the lowest laue group for the crystal system will be chosen as the symmetry of the data set. If you know that the symmetry is higher than the minimum, or if you want to collect as many Friedel pairs as possible, you can change the point group here.

When lowering the symmetry using this parameter, please remember that some (many?) Friedel pairs will always be measured, but unless the pointgroup is acentric, that will not be one of the goals of the data collection strategy. Please note that switching to an acentric pointgroup will make the data collection strategy go a long way to make sure that all Friedel pairs are measured. This might result in a much less efficient strategy (collecting the full sphere is much more than 2 times more difficult than collecting a half sphere). Remember that the absolute configuration of the structure is probably apparent from a few low-resolution reflections, and hence one does not need a complete set of Friedel pairs to determine the absolute configuration. For borderline cases (very weak anomalous scattering), please also see the description of the Friedel strategy (below).

Coinciding axes criterion
This is a value used by Y.Le Page's algorithm to determine the symmetry of the lattice. If you have an inaccurate cell, this value might need to be raised above the default of 0.2 degrees, in case of pseudo-symmetry it might need to be lowered.
Should unit cell be normalized
If this is switched to "No", the Buerger cell reduction and Le Page lattice symmetry algorithms will not be used. The crystal system will be guessed from the given unit cell parameters using the "Coinciding axes criterion".

Note: If a ".rmat" file includes a transformation matrix, that transformation is used to determine the conventional cell instead of using the Y.Le Page algorithm (See the rmat format specifications)

Press the "OK" button to continue. Yet another dialog box will pop up, requesting the host name of the Windows PC that controls the goniometer.

Hardware parameters window

CCD Controller host name
If you have only one goniometer, this name should be correct, and all that is required is to press the button labelled "OK" to make the connection.

If your generator is remote-accessible, and the settings are lower than "collect" thinks they should be (to configure what collect thinks is too low, please see configuration of generator settings), you will now get a one-time warning.

Generator settings too low?

If your generator is in "REMOTE" mode, you can change the generator setting later when making the "scan set" if desired.

At this point the main experiment window will show the current status: the hardware connection, the unit cell, and the calculated dataset.

Collect with datacollection experiment, before activating strategy

For the hardware connection, the following buttons may be present:

For the cell, a representation of the cell parameters is shown in three lines. The first line is the cell as read from the input file, the second line is a Buerger-reduced cell (Z. Kristall. 113 (1960) pp 52--56), and the third line is a standard cell as found by the algorithm of Y. Le Page (J. Appl. Cryst. 15 (1982) 255--259). The automatically determined crystal system is shown on the third line as well. If the cell standardization was inhibited, the first and last cells will be identical. Press the button to change the cell.

For the data set, a line will be shown summarizing the hkl limits and the internal symmetry. Press the button to change the parameters.

At this point, an empty scanset will appear in a separate window, plus a number of buttons that enable you to change the scanset.

Empty scan set for data collection

The most important button is the one labeled "Strategy". If the strategy button is pressed, a dialog box containing the parameters of the strategy will be shown:

Data collection strategy parameters

Those are:

Detector to crystal distance
A strategy is normally carried out at a single detector to crystal distance. The default value will be calculated from earlier results (if available), if nothing is known yet, it is set to the value of a configuration parameter. The distance needs to be large enough to be able to separate the individual diffraction spots (For Mo K-alpha radiation this can normally be accomplished if the detector to crystal distance in mm is larger than the the longest axis of the unit cell in Angstrom). Obviously, if this value is set to unnecessarily large values, the measurement will take longer than needed.
Datacollection strategy
Choose the type of strategy here. Normally, a phi scan is made that captures as many reflections as possible, followed by omega scans to complete the dataset. Alternative strategies can be selected using the arrow at the right of the entry. There are a number of specialized strategies for special cases (described below). If you prefer one of the alternative strategies as a default, this can be set using a configuration variable
Enable low-temperature kappa limitations
For low-temperature measurements, |kappa| should be restricted to values between 15 and 120 degrees. At |kappa| lower than 15 degrees excessive cooling of the phi motor could occur, and at |kappa| higher than 120 degrees icing of the crystal could result. Enable this option to enforce these kappa limitations.
Under "More options", there are more options for more unusual experiments:

Data collection strategy parameters; expert view
Prevent overlap by the longest axis
If the longest axis in the reduced cell is very long, normally very fine slices must be taken to prevent overlap of neighbouring reflections. Switching this flag to "on" tries to avoid this type of overlap by avoiding to point the long vector in the direction of the detector, allowing for bigger slices. This is especially useful if the longest axis of the reduced cell is much larger than the other two.
Enable strict efficiency theta limitations
Normally one does not allow the detector swing angle to be higher than strictly necessary for the outside edge to reach the highest reflections that should be collected. However, when collecting the full sphere of data, collection of some high-resolution reflections might require a larger detector-swing. In such cases this limitation can be removed. Note, however, that the integration of high-swing scans might be problematic, and that datacollection efficiency will suffer. Please remember that there are normally better ways of obtaining a redundant data set than to enforce a full sphere. The determination of absolute configuration is performed best by accurately collecting redundant low-resolution data (e.g. using a Friedel strategy) and not by measuring as many Friedel pairs as possible.
Desired completeness
The strategy calculation will stop when the data collection is sufficiently complete. The default completeness, 0.995, should normally be fine.
Desired redundancy of the dataset
If this is set to a value larger than 1.0, the strategy will try to collect most reflections (normally 90%) at least that number of times. This can be useful if absorption correction is required or if extremely accurate data are needed for e.g. electron density studies.
The strategies available are:
Make phi scans + omega scans to fill asymmetric unit
Standard catch-all strategy. Choose this one unless you have very good reasons to choose another.
Make optimized phi scans at kappa=0
Skips the omega scans, only makes phi scans. This is not guaranteed to give a complete data set.
Make omega scans to fill asymmetric unit
Skips the phi scans, only makes omega scans.
Make omega scans with even order axes vertical + other omega scans
Tries to measure as much as possible using 2 fold axes in the lattice vertical. The rest of the data collection is done using other omega scans. This is also known as the "Friedel" strategy, since for all lattice axes that are also axes in the structure, Friedel pairs will be measured in the same frame minimizing errors in the Bijvoet differences and optimizing the discriminating power between the enantiomeric structures.
Make a phi scan + 2 standard omega scans
This strategy (at dx=35 mm) can collect a half sphere. It has relatively low redundancy, but is very quick to calculate, and makes "the same" measurements regardless of the crystal.
Make 360 degree phi scans at kappa=0
For when you want that. This is not guaranteed to give a complete data set.
Make 180 degree phi scans at kappa=0
For when you want that. This is not guaranteed to give a complete data set.
Make omega scans at chi=55 degrees
A quick and easy strategy using only scans at chi=+/-55 degrees, at 120 degree intervals of phi. Not guaranteed to give a complete data set.
Suggestions for other data collection strategies are always welcome.

After selecting the right strategy, press "OK" to proceed. All necessary information has now been supplied, and the strategy will be calculated.

Calculating strategy

Depending on the complexity of the question this can take anywhere between a few seconds and several minutes to complete. During this time, "collect" talks to your goniometer controller to ask it about hardware limitations.

The calculated strategy will be displayed in the strategy editor. At the right side in a yellow background the number of "new" reflections measured by each scan is shown.

Finished data collection strategy

You can (temporarily) remove scans from the strategy using the checkboxes at the beginning of each line. A permanent removal can be achieved by pressing the "X" button to the right of each scan, or by leaving the scan set editor using "OK" while scans are deselected.

The "Overflow scans" button in the scan set editor achieves the same as the strategy for a subset of low-resolution reflections. It will add the necessary scans to the end of the strategy, and will make sure that they are measured at a much higher speed.

Scans can be edited by clicking on the line itself using the left mouse button:

Editing a phi scan

Some expert options are hiding under the "More options" button:

Editing a phi scan; expert view

The extra two lines allow the user to require that a specific scan uses a different data collection time or a different frame angle than all other scans. This can be specified by "overwriting" the normal values or by "multiplying" the normal values by a factor. These extra parameters can be useful to make a quick data collection to capture overloaded reflections at low resolutions.

Click "Ok" to continue (and to confirm any changes you may have made).

Back in the scan set window, the button "Remove deselected" removes from the display those scans that have their checkboxes disabled. The "Unique Data" button can be used to calculate the effect of the changes on the completeness of the measured data set.

Pressing the "Redundancy" button will start an alternative procedure that will calculate the redundancy of the dataset after each scan, and the number of reflections with the lowest redundancy:

Redundancy calculations

In this example, a total of 3525 reflections will be measured. 90% of all data will be measured 1.0 times or more, and 104 reflections were measured only once (and no reflections were missed).

A more complete overview of the redundancy of the final dataset is shown in another window that will pop up:

Redundancy overview

Click "Ok" to dismiss this window and continue building the scan set.

Using the "Add operation" button, you can manually add scans, generator settings, cryostat commands, and dark images to the data collection. The defaults for the generator settings, cryostat control and view position can be set in the collect configuration files.

Add operation

This will be mainly used to change the generator settings before and/or after the measurement:

Complete scan set program

We have already seen that the individual operations in the scan set can be modified by clicking the left mouse button on the green area. Other manipulations (moving the operations around or duplicating them) is possible by clicking the right mouse button over the operation description. A small menu will pop up from which the desired operation can be selected.

Press "OK" to complete the generation of the scan set. You will notice that the "Make scan-set" button has now turned green, indicating that a valid (non-empty) scan set has been constructed.

Now a Dialog window pops up asking for the scan parameters.

Scan parameters for data collection

Under the "More options" button, you will find:

Scan parameters for data collection; expert view

On pressing "Apply", the expected end time (normally accurate to within about 15 minutes) will be updated to represent the new parameters. The "Change" button can be used to change the integration time such that the measurement is finished e.g. just before 17.00 today or just before 9.00 tomorrow morning.

Click "OK".

At this point it is desirable to save the experiment. Select "Save experiment" from the "File" menu, and give the experiment a name (If the experiment is not saved, "collect" will attempt to auto-save the experiment as "autosave.non").

For advanced users: it is possible to change the unit cell (e.g. make it triclinic, or lowering the symmetry) and the data set (e.g. to measure to higher resolution) using the buttons in the main experiment window, and re-enter "Edit scan set". You can then complete the new data set by calculating another strategy. The target dataset for the additional strategy is the dataset listed in the status row at the bottom of the scan set window.

An example:

Measuring Friedel pairs

Bijvoet differences are most notable in low-resolution reflections. It may be inefficient to measure all Friedel pairs out to high diffraction angles. Instead:
  • First make a dataset upto 27.5 degrees theta, and use it to make a base-strategy. Press OK on the scan-set window to return to the main window of collect.
  • Click on the unit cell line, and change to a non-centric point group. Chances are that there are multiple possibilities. Choose the right one.
  • Now click on the dataset line, and change "theta max" to a lower angle, like 15 degrees. OK.
  • Re-enter "edit scan set", Use "Strategy" to find an efficient way of measuring the additional data. If you expect very small Bijvoet differences, you can use the Friedel strategy.

If you press "Collect Data" in the main window, the measurement will be started. If possible, the disk-space available on the current disk is checked just before.

Not enough disk space

or even:

By far not enough disk space

Obviously, this warning will not appear if you have enough diskspace when the measurement starts, but subsequently fill up your disk while doing other projects. If the disk fills up, data collection will be stopped.

If, for some reason, the measurement would stop or be stopped in the middle, simply press the "Collect data" button again to continue. All images that have been collected so far will automatically be skipped. Do not remove e.g. scan 1 from the scan set before clicking "Collect data" again, because this will effectively renumber all scans, and what was earlier meant to be scan 2 will now be collected as "s01f###.kcd"!


index previous next

Previous: collect: Data collection strategy and data collection
Next: Cell determination using Denzo


(C) 1997-2008, Bruker AXS BV, R.W.W. Hooft