autocal: Automatically calibrate a KappaCCD

The "autocal" program can be used to go through all the necessary steps of a KappaCCD calibration in an automated way. On a partly calibrated system, only the steps necessary to complete the calibration will be performed.

For a description of all the steps in the calibration see the manuals of the individual calibration tools and the description of the step-wise calibration procedure

The program will progress without asking any questions for as long as it can. When human interaction is required, it will start an irritating beeping sequence until the issue is resolved. If the program is not absolutely sure that everything went 100% OK, it will ask you to consult an expert. In most of these cases nothing serious will be wrong, but it is in any case very valuable to send such messages to collect@bruker-axs.nl such that we can check, and possibly improve the calibration software to handle your case more smoothly.

If everything proceeds successfully, the complete calibration from scratch can be performed in 24 hours. To reach this, the calibration process should be started in the morning. The last human interaction (a sample change) should be required approximately 6 hours after the start.

The goal of the autocal program is to perform only those calibrations that have not been performed before. This means that:

Prerequisites

For running autocal, one needs: If the machine has never been calibrated for collect/evalccd before, (machines coming from the factory before January 2000 were not calibrated for collect/evalccd) one additionally needs:

Usage

The program is invoked by the command line:
  autocal [options]
Options are:
noambi
Inhibit the measurement and integration of a standard AMBI data set to verify the calibration.
ambi
Enable the measurement and integration of a standard AMBI data set to verify the calibration. This is the default.
denzo
Generate calibration files in a form suitable for the HKL2000 system as well. This is the default.
nodenzo
Do not generate calibration files in a form suitable for the HKL2000 system.
factory
Generate some extra images only needed in a factory calibration. This is the default on machines named "nows65".
nofactory
Do not generate extra images only needed in a factory calibration. This is the default on machines not named "nows65".
mkdir
With this option, autocal will not perform any calibration measurements but only check whether the calibration directories exist, and if not, create them. The program will then show the names of these directories and exit.